CLASS II Device recall · Reported 29 Mar 2017 · FDA Enforcement Report

Giraffe Bedded Warmer, Panda Warmer Modification to Giraffe recalled

Ohmeda Medical is recalling Giraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe, distributed nationwide in the US. The recall was initiated on 18 Nov 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSee Attached Spreadsheat
Quantity recalled5,993 (5,080 US; 913 OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1507-2017
FDA event ID
76283
Recalling firm
Ohmeda Medical, Laurel, MD
Classification
Class II
Status
Terminated
Recall initiated
18 Nov 2016
FDA classified
20 Mar 2017
Reported
29 Mar 2017
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

18 Nov 2016Recall initiated by company
20 Mar 2017Classified by FDA+122 days
29 Mar 2017Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Giraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe and Panda Uninterruptible Power Supply) - Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Where it was distributed

Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Which lots are affected?

See Attached Spreadsheat

Where was it sold?

Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

10 other products · FDA event 76283

This product is part of a recall event; the main page for the event is Giraffe Bedded Warmer, Panda Warmer Giraffe recalled by Ohmeda Medical.

More recalls from Ohmeda Medical

All 13

Other Patient Monitors recalls

All 1,013