CLASS II ONGOING Device recall · Reported 7 May 2025 · FDA Enforcement Report

IceSeed 1.5 CX S Needle US, Cryoablation Needle recalled

Boston Scientific is recalling IceSeed 1.5 CX S Needle US, Cryoablation Needle, distributed nationwide in the US. The recall was initiated on 24 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN 00191506032678, Batch Numbers: 33744602, 33744603, 33744604, 33744605, 33931901, 33937786, 33937787, 33942987, 33942988, 33942989, 33950417, 33950418, 33996423, 33996424, 33996428, 34038291, 34038292, 34038307, 34117281, 34128141, 34128142, 34151877, 34151878, 34204928, 34204929, 34256930, 34256931, 34256932, 34295233.
Quantity recalled435 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1686-2025
FDA event ID
96556
Recalling firm
Boston Scientific, Maple Grove, MN
Classification
Class II
Status
Ongoing
Recall initiated
24 Mar 2025
FDA classified
29 Apr 2025
Reported
7 May 2025
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Temperature Abuse

Timeline

24 Mar 2025Recall initiated by company
29 Apr 2025Classified by FDA+36 days
7 May 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

Which lots are affected?

GTIN 00191506032678, Batch Numbers: 33744602, 33744603, 33744604, 33744605, 33931901, 33937786, 33937787, 33942987, 33942988, 33942989, 33950417, 33950418, 33996423, 33996424, 33996428, 34038291, 34038292, 34038307, 34117281, 34128141, 34128142, 34151877, 34151878, 34204928, 34204929, 34256930, 34256931, 34256932, 34295233.

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 96556

This product is part of a recall event; the main page for the event is IceSeed 1.5 CX S Needle OUS, Cryoablation Needle recalled.

More recalls from Boston Scientific

All 628

Other Syringes & Needles recalls

All 1,066