CLASS II Device recall · Reported 22 Jan 2014 · FDA Enforcement Report

Imagecast PACS is an image management system intended to be recalled

Ge Healthcare It is recalling Imagecast PACS is an image management system intended to be used by trained professionals, distributed nationwide in the US. The recall was initiated on 4 Mar 2011 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesImagecast PACS versions prior to 3.6.51.01
Quantity recalled43 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0773-2014
FDA event ID
64740
Recalling firm
Ge Healthcare It, Barrington, IL
Classification
Class II
Status
Terminated
Recall initiated
4 Mar 2011
FDA classified
15 Jan 2014
Reported
22 Jan 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

4 Mar 2011Recall initiated by company
15 Jan 2014Classified by FDA+1048 days
22 Jan 2014Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas

Product as listed by the FDA

Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.

Where it was distributed

Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.

Nationwide AlabamaArizonaConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIndianaLouisianaMassachusettsMichiganMinnesotaMissouriNorth CarolinaNew HampshireNew JerseyNew YorkOhioPennsylvaniaTexas

Questions about this recall

Why was it recalled?

In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas

Which lots are affected?

Imagecast PACS versions prior to 3.6.51.01

Where was it sold?

Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.