CLASS II Device recall · Reported 1 Sep 2021 · FDA Enforcement Report

Ingenia 0T-Magnetic Resonance Medical Electrical Systems recalled

Philips North America is recalling Ingenia 0T-Magnetic Resonance Medical Electrical Systems indicated for use as a, distributed nationwide in the US. The recall was initiated on 30 Jul 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesS/N and UDI: 85275 (01)00884838055339(21)85275; 85278 (01)00884838055339(21)85278; 85394 (01)00884838055339(21)85394. OUS: 85393 (01)00884838055339 85332 (01)00884838055339 85334 (01)00884838055339 85336 (01)00884838055339 85349 (01)00884838055339 85350 (01)00884838055339 85355 (01)00884838055339 85370 (01)00884838055339 85371 (01)00884838055339 85382 (01)00884838055339 85384 (01)00884838055339 85385 (01)00884838055339 85398 (01)00884838055339 85348 (01)00884838055339 85149 (01)00884838055339 85367 (01)00884838055339 85388 (01)00884838055339
Quantity recalled3 units US; 17 units OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2343-2021
FDA event ID
88461
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Terminated
Recall initiated
30 Jul 2021
FDA classified
25 Aug 2021
Reported
1 Sep 2021
Type
Voluntary: Firm initiated
Customers notified by
N/A
Reason category
Other Reasons

Timeline

30 Jul 2021Recall initiated by company
25 Aug 2021Classified by FDA+26 days
1 Sep 2021Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Product as listed by the FDA

Ingenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781342

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoMarylandNew HampshireNew YorkOhioTennesseeTexasVirginia

Questions about this recall

Is Ingenia 0T-Magnetic Resonance Medical Electrical Systems indicated for use as a recalled?

Yes. Philips North America recalled Ingenia 0T-Magnetic Resonance Medical Electrical Systems indicated for use as a on 30 Jul 2021. The recall is Class II and its status is Terminated. Recall number Z-2343-2021.

Why was it recalled?

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Which lots are affected?

S/N and UDI: 85275 (01)00884838055339(21)85275; 85278 (01)00884838055339(21)85278; 85394 (01)00884838055339(21)85394. OUS: 85393 (01)00884838055339 85332 (01)00884838055339 85334 (01)00884838055339 85336 (01)00884838055339 85349 (01)00884838055339 85350 (01)00884838055339 85355 (01)00884838055339 85370 (01)00884838055339 85371 (01)00884838055339 85382 (01)00884838055339 85384 (01)00884838055339 85385 (01)00884838055339 85398 (01)00884838055339 85348 (01)00884838055339 85149 (01)00884838055339 85367 (01)00884838055339 85388 (01)00884838055339

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips North America

All 531

Other Test Kits & Diagnostics recalls

All 2,656