CLASS II Device recall · Reported 12 Oct 2016 · FDA Enforcement Report

Ingenuity Core recalled by Philips Medical Systems

Philips Medical Systems is recalling Ingenuity Core, distributed nationwide in the US. The recall was initiated on 24 Aug 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes30041, 31001, 33084, 52000, 52004, 52007, 52008, 52010, 52012, 52013, 52014, 52015, 52018, 52019, 52020, 52021, 52022, 52024, 52025, 52026, 52027, 52029, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52040, 52041, 52042, 52043, 52044, 52045, 52046, 52047, 52048, 52050, 25052, 52053, 52054, 52055, 52056, 52057, 52058, 52059, 52061, 52063, 52064, 52065, 52067, 52068, 52069, 52070, 52071, 52072, 52073, 52074, 52075, 52076, 52077, 52078, 52079, 52080, 52081, 52082, 52083, 52084, 132164, 132465, 310001, 310002, 310003, 310009, 310017, 310019, 310022, 310023, 310026, 310032, 310039, 310042, 310052, 310054, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310090, 310096, 310098, 310105, 310107, 310111, 310113, 310116, 310119, 310121, 310125, 310129, 310133, 310134, 310135, 310141, 310143, 310149, 310150, 310151, 310157, 310159, 310160, 310170, 310203, 310204, 310205, 310206, 310207, 310208, 310209, 310210, 310211, 310212, 310213, 310214, 310215, 310216, 310217, 310218, 310219, 310220, 310221, 310222, 310223, 310224, 310225, 310226, 310227, 310228, 310229, 310230, 310231, 310232, 310233, 310234, 310235, 310236, 310237, 310238, 310239, 310240, 310241, 310242, 310243, 310244, 310245, 310246, 310247, 310248, 310249, 310250, 310251, 310252, 310253, 310254, 310255, 310256, 310257, 310258, 310259, 310260, 310261, 310262, 310263, 310264, 310265, 310266, 310267, 310268, 310269, 310270, 310271, 310272, 310273, 310274, 310275, 310276, 310277, 310278, 310279, 310280, 310281, 310282, 310283, 310284, 310285, 310286, 310287, 310288, 310289, 310290, 310291, 310292, 310293, 310294, 310295, 310296, 310297, 310298, 310299, 310300, 310301, 310302, 310303, 310304, 310305, 310306, 310307, 310308, 310309, 310310, 310311, 310313, 310315, 310316, 310317, 310318, 310319, 310320, 310321, 310322, 310323, 310324, 310325, 310326, 310327, 310328, 310329, 310330, 310331, 333002, 333005, 333009, 333010, 333014, 333015, 333018, 333019, 333021, 333030, 333031, 333037, 333041, 333042, 333043, 333044, 333045, 333046, 333048, 333049, 333050, 333051, 333052, 333053, 333054, 333055, 333056, 333059, 333060, 333061, 333062, 333063, 333064, 333065, 333066, 333067, 333068, 333069, 333070, 333071, 333072, 333073, 333074, 333075, 333076, 333077, 333078, 333079, 333080, 333081, 333082, 333083, 333085, 333086, 333087, 333088, 333089, 333090, 333091, 333092, 333093, 333094, 333095, 333096, 333097, 333098, 333099, 333100, 333101, 333102, 333103, 333104, 333105, 333106, 333107, 333108, 333109, 333110, 333111, 333112, 333113, 333114, 333115, 333116, 333117, 333118, 333119, 333120, 333121, 333122, 333123, 333124, 333125, 333126, 333127, 333129, 333130, 333131, 333132, 333133, 333134, 333135, 333136, 333137, 333140, 333144, 333145, 333146, 333150, 333151
Quantity recalled358

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0010-2017
FDA event ID
75157
Recalling firm
Philips Medical Systems, Cleveland, OH
Classification
Class II
Status
Terminated
Recall initiated
24 Aug 2016
FDA classified
3 Oct 2016
Reported
12 Oct 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

24 Aug 2016Recall initiated by company
3 Oct 2016Classified by FDA+40 days
12 Oct 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

Product as listed by the FDA

Ingenuity Core Model Number 728321 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Where it was distributed

Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichigan

Questions about this recall

Is Ingenuity Core recalled?

Yes. Philips Medical Systems recalled Ingenuity Core on 24 Aug 2016. The recall is Class II and its status is Terminated. Recall number Z-0010-2017.

Why was it recalled?

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

Which lots are affected?

30041, 31001, 33084, 52000, 52004, 52007, 52008, 52010, 52012, 52013, 52014, 52015, 52018, 52019, 52020, 52021, 52022, 52024, 52025, 52026, 52027, 52029, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52040, 52041, 52042, 52043, 52044, 52045, 52046, 52047, 52048, 52050, 25052, 52053, 52054, 52055, 52056, 52057, 52058, 52059, 52061, 52063, 52064, 52065, 52067, 52068, 52069, 52070, 52071, 52072, 52073, 52074, 52075, 52076, 52077, 52078, 52079, 52080, 52081, 52082, 52083, 52084, 132164, 132465, 310001, 310002, 310003, 310009, 310017, 310019, 310022, 310023, 310026, 310032, 310039, 310042, 310052, 310054, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310090, 310096, 310098, 310105, 310107, 310111, 310113, 310116, 310119, 310121, 310125, 310129, 310133, 310134, 310135, 310141, 310143, 310149, 310150, 310151, 310157, 310159, 310160, 310170, 310203, 310204, 310205, 310206, 310207, 310208, 310209, 310210, 310211, 310212, 310213, 310214, 310215, 310216, 310217, 310218, 310219, 310220, 310221, 310222, 310223, 310224, 310225, 310226, 310227, 310228, 310229, 310230, 310231, 310232, 310233, 310234, 310235, 310236, 310237, 310238, 310239, 310240, 310241, 310242, 310243, 310244, 310245, 310246, 310247, 310248, 310249, 310250, 310251, 310252, 310253, 310254, 310255, 310256, 310257, 310258, 310259, 310260, 310261, 310262, 310263, 310264, 310265, 310266, 310267, 310268, 310269, 310270, 310271, 310272, 310273, 310274, 310275, 310276, 310277, 310278, 310279, 310280, 310281, 310282, 310283, 310284, 310285, 310286, 310287, 310288, 310289, 310290, 310291, 310292, 310293, 310294, 310295, 310296, 310297, 310298, 310299, 310300, 310301, 310302, 310303, 310304, 310305, 310306, 310307, 310308, 310309, 310310, 310311, 310313, 310315, 310316, 310317, 310318, 310319, 310320, 310321, 310322, 310323, 310324, 310325, 310326, 310327, 310328, 310329, 310330, 310331, 333002, 333005, 333009, 333010, 333014, 333015, 333018, 333019, 333021, 333030, 333031, 333037, 333041, 333042, 333043, 333044, 333045, 333046, 333048, 333049, 333050, 333051, 333052, 333053, 333054, 333055, 333056, 333059, 333060, 333061, 333062, 333063, 333064, 333065, 333066, 333067, 333068, 333069, 333070, 333071, 333072, 333073, 333074, 333075, 333076, 333077, 333078, 333079, 333080, 333081, 333082, 333083, 333085, 333086, 333087, 333088, 333089, 333090, 333091, 333092, 333093, 333094, 333095, 333096, 333097, 333098, 333099, 333100, 333101, 333102, 333103, 333104, 333105, 333106, 333107, 333108, 333109, 333110, 333111, 333112, 333113, 333114, 333115, 333116, 333117, 333118, 333119, 333120, 333121, 333122, 333123, 333124, 333125, 333126, 333127, 333129, 333130, 333131, 333132, 333133, 333134, 333135, 333136, 333137, 333140, 333144, 333145, 333146, 333150, 333151

Where was it sold?

Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.