CLASS I Device recall · Reported 31 Aug 2016 · FDA Enforcement Report

INRatio Prothrombin Time (PT) Monitoring System recalled

Alere San Diego is recalling INRatio Prothrombin Time (PT) Monitoring System, distributed nationwide in the US. The recall was initiated on 11 Jul 2016 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesAll Lots All Serial Numbers

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2355-2016
FDA event ID
74665
Recalling firm
Alere San Diego, San Diego, CA
Classification
Class I
Status
Terminated
Recall initiated
11 Jul 2016
FDA classified
19 Aug 2016
Reported
31 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
Press Release
Reason category
Other Reasons

Timeline

11 Jul 2016Recall initiated by company
19 Aug 2016Classified by FDA+39 days
31 Aug 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Product as listed by the FDA

INRatio Prothrombin Time (PT) Monitoring System Model Number: 0100007 Product Usage: The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test strip area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the Prothrombin time test on the blood sample. The other two channels contain the reagents. to perfom the Prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control test are required for the INRatio System. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result.

Where it was distributed

Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.

Nationwide

Questions about this recall

Why was it recalled?

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Which lots are affected?

All Lots All Serial Numbers

Where was it sold?

Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

8 other products · FDA event 74665

This product is part of a recall event; the main page for the event is Alere INRatio PT/INR Test Strips recalled by Alere San Diego.

More recalls from Alere San Diego

All 54

Other Insulin Pumps & Glucose Monitors recalls

All 467