CLASS II ONGOING Device recall · Reported 22 Feb 2023 · FDA Enforcement Report
iTotal Identity Impactor Handle recalled by Conformis
Conformis is recalling iTotal Identity Impactor Handle, distributed nationwide in the US. The recall was initiated on 14 Nov 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1133-2023
- FDA event ID
- 91376
- Recalling firm
- Conformis, Billerica, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 14 Nov 2022
- FDA classified
- 13 Feb 2023
- Reported
- 22 Feb 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 14 Nov 2022 | Recall initiated by company | |
| 13 Feb 2023 | Classified by FDA | +91 days |
| 22 Feb 2023 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may loosen and fall into the patient during a procedure.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
iTotal Identity Impactor Handle, Model No. 1080-212/ED-07827 REV AE
Where it was distributed
US Nationwide distribution in the states of AL, AZ, CO, FL, GA, KS, KY, LA, MA, MD, MS, NE, NH, NV, NY, OH, TN, WI, WY.
Nationwide AlabamaArizonaColoradoFloridaGeorgiaKansasKentuckyLouisianaMassachusettsMarylandMississippiNebraskaNew HampshireNevadaNew YorkOhioTennesseeWisconsinWyoming
Questions about this recall
Why was it recalled?
Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may loosen and fall into the patient during a procedure.
Which lots are affected?
UDI-DI 00810933030582 Lot D221301
Where was it sold?
US Nationwide distribution in the states of AL, AZ, CO, FL, GA, KS, KY, LA, MA, MD, MS, NE, NH, NV, NY, OH, TN, WI, WY.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Conformis
All 34| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | restor3d Cordera Hip System recalled by Conformis | Conformis | Labeling Errors | Nationwide |
| Device | CLASS II | Identity Imprint PS Tibial Tray Size recalled by Conformis | Conformis | Packaging Defects | Nationwide |
| Device | CLASS II | Hip-screw-, Sterile Eo recalled by Conformis | Conformis | Other | Nationwide |
| Device | CLASS II | iTotal Identity Posterior Stabilised (PS) Knee Replacement System recalled | Conformis | Packaging Defects | MS |
| Device | CLASS II | iTotal Hip Replacement System recalled by Conformis | Conformis | Other | CA |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |