CLASS II Device recall · Reported 11 Apr 2018 · FDA Enforcement Report
LeadCare II Blood Lead Analyzer Part Numbers recalled
Magellan Diagnostics is recalling LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. The recall was initiated on 9 Jan 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1295-2018
- FDA event ID
- 79279
- Recalling firm
- Magellan Diagnostics, North Billerica, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 Jan 2018
- FDA classified
- 3 Apr 2018
- Reported
- 11 Apr 2018
- Reason category
- Other Reasons
- Device type
- Test Kits & Diagnostics
Timeline
| 9 Jan 2018 | Recall initiated by company | |
| 3 Apr 2018 | Classified by FDA | +84 days |
| 11 Apr 2018 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Product as listed by the FDA
LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
Where it was distributed
Worldwide Distribution
Questions about this recall
Why was it recalled?
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Which lots are affected?
Part Numbers: 70-6760, 70-6529-R, 70-6760-P, and 70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.
Where was it sold?
Worldwide Distribution
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 79279| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | LeadCare Plus Blood Lead Analyzer recalled by Magellan Diagnostics | Magellan Diagnostics | Other |
More recalls from Magellan Diagnostics
All 17| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | LeadCare II Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Other | Nationwide |
| Device | CLASS II | LeadCare Ultra Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | (2) LeadCare¿ II Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Packaging Defects | Nationwide |
| Device | CLASS II | LeadCare¿ Plus Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Packaging Defects | Nationwide |
| Device | CLASS II | The LeadCare¿ II Blood Lead Test System relies recalled | Magellan Diagnostics | Labeling Errors | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,675| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Nextseq 550Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | NovaSeq 6000Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | Liaison Plex Gram-Negative Blood Culture Assay recalled by Luminex | Luminex | Potency & Assay | Nationwide |
| Device | CLASS II | MiSeq Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | Techno-Path Multichem AE Unassayed Level recalled | Techno-path Manufacturing | Potency & Assay | Nationwide |