CLASS II Device recall · Reported 11 Apr 2018 · FDA Enforcement Report

LeadCare II Blood Lead Analyzer Part Numbers recalled

Magellan Diagnostics is recalling LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. The recall was initiated on 9 Jan 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Numbers: 70-6760, 70-6529-R, 70-6760-P, and 70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.
Quantity recalled6,143

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1295-2018
FDA event ID
79279
Recalling firm
Magellan Diagnostics, North Billerica, MA
Classification
Class II
Status
Terminated
Recall initiated
9 Jan 2018
FDA classified
3 Apr 2018
Reported
11 Apr 2018
Reason category
Other Reasons

Timeline

9 Jan 2018Recall initiated by company
3 Apr 2018Classified by FDA+84 days
11 Apr 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.

Product as listed by the FDA

LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.

Where it was distributed

Worldwide Distribution

Questions about this recall

Why was it recalled?

The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.

Which lots are affected?

Part Numbers: 70-6760, 70-6529-R, 70-6760-P, and 70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.

Where was it sold?

Worldwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 79279
ReportedClassRecallCompanyReasonWhere
CLASS IILeadCare Plus Blood Lead Analyzer recalled by Magellan DiagnosticsMagellan DiagnosticsOther

More recalls from Magellan Diagnostics

All 17

Other Test Kits & Diagnostics recalls

All 2,675
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IINextseq 550Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IINovaSeq 6000Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IILiaison Plex Gram-Negative Blood Culture Assay recalled by LuminexLuminexPotency & AssayNationwide
DeviceCLASS IIMiSeq Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IITechno-Path Multichem AE Unassayed Level recalledTechno-path ManufacturingPotency & AssayNationwide