CLASS I Device recall · Reported 24 May 2017 · FDA Enforcement Report

LeadCare Ultra Blood Analyzer recalled by Magellan Diagnostics

Magellan Diagnostics is recalling LeadCare Ultra Blood Analyzer, distributed nationwide in the US. The recall was initiated on 29 Nov 2014 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesAll serial numbers
Quantity recalled49 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1925-2017
FDA event ID
76991
Recalling firm
Magellan Diagnostics, North Billerica, MA
Classification
Class I
Status
Terminated
Recall initiated
29 Nov 2014
FDA classified
17 May 2017
Reported
24 May 2017
Reason category
Foreign Material

Timeline

29 Nov 2014Recall initiated by company
17 May 2017Classified by FDA+900 days
24 May 2017Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Which lots are affected?

All serial numbers

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 76991

This product is part of a recall event; the main page for the event is LeadCare Ultra Blood Lead Test Kits recalled by Magellan Diagnostics.

More recalls from Magellan Diagnostics

All 17

Other Test Kits & Diagnostics recalls

All 2,675
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IINextseq 550Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IINovaSeq 6000Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IILiaison Plex Gram-Negative Blood Culture Assay recalled by LuminexLuminexPotency & AssayNationwide
DeviceCLASS IIMiSeq Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IITechno-Path Multichem AE Unassayed Level recalledTechno-path ManufacturingPotency & AssayNationwide