CLASS I Device recall · Reported 24 May 2017 · FDA Enforcement Report
LeadCare Ultra Blood Lead Test Kits recalled by Magellan Diagnostics
Magellan Diagnostics is recalling LeadCare Ultra Blood Lead Test Kits, distributed nationwide in the US. The recall was initiated on 29 Nov 2014 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1926-2017
- FDA event ID
- 76991
- Recalling firm
- Magellan Diagnostics, North Billerica, MA
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 29 Nov 2014
- FDA classified
- 17 May 2017
- Reported
- 24 May 2017
- Reason category
- Foreign Material
- Device type
- Test Kits & Diagnostics
Timeline
| 29 Nov 2014 | Recall initiated by company | |
| 17 May 2017 | Classified by FDA | +900 days |
| 24 May 2017 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010
Where it was distributed
Nationwide Distribution
Questions about this recall
Why was it recalled?
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Which lots are affected?
Kit lots: 1309BU,1310AU,1312BU,1404AU,1408BU
Where was it sold?
Nationwide Distribution
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 76991| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | LeadCare Ultra Blood Analyzer recalled by Magellan Diagnostics | Magellan Diagnostics | Foreign Material | Nationwide |
More recalls from Magellan Diagnostics
All 17| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | LeadCare II Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Other | Nationwide |
| Device | CLASS II | LeadCare Ultra Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | (2) LeadCare¿ II Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Packaging Defects | Nationwide |
| Device | CLASS II | LeadCare¿ Plus Blood Lead Test Kit recalled by Magellan Diagnostics | Magellan Diagnostics | Packaging Defects | Nationwide |
| Device | CLASS II | The LeadCare¿ II Blood Lead Test System relies recalled | Magellan Diagnostics | Labeling Errors | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,675| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Nextseq 550Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | NovaSeq 6000Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | Liaison Plex Gram-Negative Blood Culture Assay recalled by Luminex | Luminex | Potency & Assay | Nationwide |
| Device | CLASS II | MiSeq Dx Instrument recalled by Illumina | Illumina | Other | Nationwide |
| Device | CLASS II | Techno-Path Multichem AE Unassayed Level recalled | Techno-path Manufacturing | Potency & Assay | Nationwide |