CLASS II ONGOING Device recall · Reported 30 Apr 2025 · FDA Enforcement Report

Magec 2 Rod Standard recalled by Globus Medical

Globus Medical is recalling Magec 2 Rod Standard, distributed nationwide in the US. The recall was initiated on 3 Apr 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVersion/Model: MC2-5090S; UDI-DI (with lot nuber): (01)00887517834041(11)250311(17)300311(10)GB1607ET
Quantity recalled1 unit

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1657-2025
FDA event ID
96615
Recalling firm
Globus Medical, West Norriton, PA
Classification
Class II
Status
Ongoing
Recall initiated
3 Apr 2025
FDA classified
23 Apr 2025
Reported
30 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

3 Apr 2025Recall initiated by company
23 Apr 2025Classified by FDA+20 days
30 Apr 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

Product as listed by the FDA

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

Where it was distributed

US Nationwide distribution in the state of AR.

Nationwide Arkansas

Questions about this recall

Why was it recalled?

Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

Which lots are affected?

Version/Model: MC2-5090S; UDI-DI (with lot nuber): (01)00887517834041(11)250311(17)300311(10)GB1607ET

Where was it sold?

US Nationwide distribution in the state of AR.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Globus Medical

All 29

Other Other Medical Devices recalls