CLASS I ONGOING Device recall · Reported 22 Jan 2025 · FDA Enforcement Report
Medline Fluid Delivery Set with Drip Chamber recalled
Medline Industries LP - Northfield is recalling Medline Fluid Delivery Set with Drip Chamber: FDS,135,CHMN,W/F,MLL,RC(2),PG, distributed nationwide in the US. The recall was initiated on 16 Dec 2024 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0764-2025
- FDA event ID
- 95988
- Recalling firm
- Medline Industries LP - Northfield, Northfield, IL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 16 Dec 2024
- FDA classified
- 15 Jan 2025
- Reported
- 22 Jan 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 16 Dec 2024 | Recall initiated by company | |
| 15 Jan 2025 | Classified by FDA | +30 days |
| 22 Jan 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Product as listed by the FDA
Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223
Where it was distributed
US Nationwide Distribution
Questions about this recall
Why was it recalled?
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Which lots are affected?
1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;
Where was it sold?
US Nationwide Distribution
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 95988This product is part of a recall event; the main page for the event is Medline Fluid Delivery Set with Drip Chamber recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | Medline Fluid Delivery Set with Drip Chamber recalled | Medline Industries LP - Northfield | Other | Nationwide |
More recalls from Medline Industries LP - Northfield
All 792| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Medline medical convenience kits labeled as recalled | Medline Industries LP - Northfield | Device Malfunction | 6 states |
| Device | CLASS I | Medline medical convenience kits labeled as recalled | Medline Industries LP - Northfield | Device Malfunction | 6 states |
| Device | CLASS I | Medline ReNewal Reprocessed St. Jude Medical Response recalled | Medline Industries LP - Northfield | Particulate Matter | Nationwide |
| Device | CLASS II | The Medline General Surgery Tray is customized to meet recalled | Medline Industries LP - Northfield | Sterility | Nationwide |
| Device | CLASS II | Medline Kits containing BD SureStep Foley Tray Systems recalled | Medline Industries LP - Northfield | Labeling Errors | FL, IL |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |