CLASS I ONGOING Device recall · Reported 22 Jan 2025 · FDA Enforcement Report

Medline Fluid Delivery Set with Drip Chamber recalled

Medline Industries LP - Northfield is recalling Medline Fluid Delivery Set with Drip Chamber: FDS,135,CHMN,W/F,MLL,RC(2),PG, distributed nationwide in the US. The recall was initiated on 16 Dec 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codes1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;
Quantity recalled13,600 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0764-2025
FDA event ID
95988
Recalling firm
Medline Industries LP - Northfield, Northfield, IL
Classification
Class I
Status
Ongoing
Recall initiated
16 Dec 2024
FDA classified
15 Jan 2025
Reported
22 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

16 Dec 2024Recall initiated by company
15 Jan 2025Classified by FDA+30 days
22 Jan 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Product as listed by the FDA

Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Which lots are affected?

1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 95988

This product is part of a recall event; the main page for the event is Medline Fluid Delivery Set with Drip Chamber recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IMedline Fluid Delivery Set with Drip Chamber recalledMedline Industries LP - NorthfieldOtherNationwide

More recalls from Medline Industries LP - Northfield

All 792

Other Other Medical Devices recalls

All 13,908