CLASS II ONGOING Device recall · Reported 28 May 2025 · FDA Enforcement Report
Meta-tan Lag/compression Screw Kit recalled by Smith & Nephew
Smith & Nephew is recalling Meta-tan Lag/compression Screw Kit, distributed nationwide in the US. The recall was initiated on 18 Mar 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1806-2025
- FDA event ID
- 96586
- Recalling firm
- Smith & Nephew, Memphis, TN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Mar 2025
- FDA classified
- 19 May 2025
- Reported
- 28 May 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 18 Mar 2025 | Recall initiated by company | |
| 19 May 2025 | Classified by FDA | +62 days |
| 28 May 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Product as listed by the FDA
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Where it was distributed
Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
Nationwide AlabamaArizonaCaliforniaFloridaIowaKentuckyLouisianaMassachusettsMissouriNorth CarolinaNew HampshireNew JerseyNew MexicoNevadaNew YorkOhioTennesseeTexasWashington
Questions about this recall
Why was it recalled?
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Which lots are affected?
UDI-DI: 00885556132975. Lot: 24CT81426. Expiration: 25-Mar-2034
Where was it sold?
Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Smith & Nephew
All 259| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Evos Lateral Proximal Tibia Plate 6 Hole Left Ref72463106 recalled | Smith & Nephew | Labeling Errors | Nationwide |
| Device | CLASS II | Trigen Proximal Straight Humeral Nail recalled by Smith & Nephew | Smith & Nephew | Other | DE |
| Device | CLASS II | Biosure HA Screw Threaded metallic bone fixation fastener recalled | Smith & Nephew | Labeling Errors | Nationwide |
| Device | CLASS II | Bioraptor Suture Anchors recalled by Smith & Nephew | Smith & Nephew | Sterility | Nationwide |
| Device | CLASS II | Tandem Intl Bipolar Cocr Shell/Uhmwpe Liner recalled by Smith & Nephew | Smith & Nephew | Other | TN |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |