CLASS II ONGOING Device recall · Reported 24 Jan 2024 · FDA Enforcement Report
MINC+ Benchtop Incubator, Reference recalled by Cook Medical
Cook Medical is recalling MINC+ Benchtop Incubator, Reference, distributed in California, Louisiana, Maryland, New York. The recall was initiated on 7 Dec 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0722-2024
- FDA event ID
- 93673
- Recalling firm
- Cook Medical, Bloomington, IN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Dec 2023
- FDA classified
- 12 Jan 2024
- Reported
- 24 Jan 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Temperature Abuse
- Device type
- Other Medical Devices
Timeline
| 7 Dec 2023 | Recall initiated by company | |
| 12 Jan 2024 | Classified by FDA | +36 days |
| 24 Jan 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.
Product as listed by the FDA
MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Where it was distributed
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
Questions about this recall
Why was it recalled?
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
Which lots are affected?
UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316
Where was it sold?
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cook Medical
All 53| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Zenith Alpha 2 Thoracic Endovascular Graft recalled | Cook Medical | Other | 27 states |
| Device | CLASS II | EchoTip¿ Ultra Endoscopic Ultrasound Access Needle, Echo-19, G31520 recalled | Cook Medical | Other | IN |
| Device | CLASS II | EchoTip¿ Ultra HD Endoscopic Ultrasound Access Needle recalled | Cook Medical | Other | IN |
| Device | CLASS II | Disposable Humidification Flask, Reference recalled by Cook Medical | Cook Medical | Packaging Defects | Nationwide |
| Device | CLASS III | Gunther Tulip Vena Cava Filter Set for Jugular Vein Approach recalled | Cook Medical | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |