CLASS II ONGOING Device recall · Reported 15 May 2024 · FDA Enforcement Report

Perla Tl 25d Screw recalled by Spineart

Spineart is recalling Perla Tl 25d Screw, distributed nationwide in the US. The recall was initiated on 15 Mar 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Numbers/Lot-Batch Numbers: TLF-DS 45 25-S 7-4407 TLF-DS 45 25-S 7-6207 TLF-DS 45 30-S 7-4244 TLF-DS 45 30-S 7-6226 TLF-DS 45 35-S 7-4242 TLF-DS 45 35-S 7-5470 TLF-DS 45 35-S 7-6684 TLF-DS 45 40-S 7-6232 TLF-DS 45 45-S 7-5474 TLF-DS 45 45-S 7-6874 TLF-DS 55 30-S 7-4409 TLF-DS 55 30-S 7-4695 TLF-DS 55 30-S 7-4965 TLF-DS 55 35-S 7-4413 TLF-DS 55 35-S 7-6455 TLF-DS 55 40-S 7-4415 TLF-DS 55 40-S 7-4417 TLF-DS 55 45-S 7-4411 TLF-DS 55 45-S 7-5055 TLF-DS 55 45-S 7-6087 TLF-DS 55 45-S 7-6459 TLF-DS 65 35-S 7-4405 TLF-DS 65 35-S 7-6230 TLF-DS 65 35-S 7-6461 TLF-DS 65 35-S 7-6882 TLF-DS 65 40-S 7-4248 TLF-DS 65 40-S 7-4250 TLF-DS 65 40-S 7-5480 TLF-DS 65 40-S 7-6228 TLF-DS 65 40-S 7-6554 TLF-DS 65 40-S 7-6740 TLF-DS 65 45-S 7-4582 TLF-DS 65 45-S 7-5484 TLF-DS 65 45-S 7-5808 TLF-DS 65 45-S 7-6668 TLF-DS 65 45-S 7-6742 TLF-DS 65 45-S 7-7299 TLF-DS 65 50-S 7-5488 TLF-DS 65 50-S 7-9264 UDI-DI for associated Model Numbers: TLF-DS 45 25-S 07640185345799 TLF-DS 45 30-S 07640185345805 TLF-DS 45 35-S 07640185345812 TLF-DS 45 40-S 07640185345829 TLF-DS 45 45-S 07640185345836 TLF-DS 55 30-S 07640185345850 TLF-DS 55 35-S 07640185345867 TLF-DS 55 40-S 07640185345874 TLF-DS 55 45-S 07640185345881 TLF-DS 60 35-S 07640305166082
Quantity recalled242 systems

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1758-2024
FDA event ID
94257
Recalling firm
Spineart, Plan-Les-Ouates, N/A
Classification
Class II
Status
Ongoing
Recall initiated
15 Mar 2024
FDA classified
3 May 2024
Reported
15 May 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

15 Mar 2024Recall initiated by company
3 May 2024Classified by FDA+49 days
15 May 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Product as listed by the FDA

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).

Where it was distributed

U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.

Nationwide CaliforniaKentucky

Questions about this recall

Why was it recalled?

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Which lots are affected?

Model Numbers/Lot-Batch Numbers: TLF-DS 45 25-S 7-4407 TLF-DS 45 25-S 7-6207 TLF-DS 45 30-S 7-4244 TLF-DS 45 30-S 7-6226 TLF-DS 45 35-S 7-4242 TLF-DS 45 35-S 7-5470 TLF-DS 45 35-S 7-6684 TLF-DS 45 40-S 7-6232 TLF-DS 45 45-S 7-5474 TLF-DS 45 45-S 7-6874 TLF-DS 55 30-S 7-4409 TLF-DS 55 30-S 7-4695 TLF-DS 55 30-S 7-4965 TLF-DS 55 35-S 7-4413 TLF-DS 55 35-S 7-6455 TLF-DS 55 40-S 7-4415 TLF-DS 55 40-S 7-4417 TLF-DS 55 45-S 7-4411 TLF-DS 55 45-S 7-5055 TLF-DS 55 45-S 7-6087 TLF-DS 55 45-S 7-6459 TLF-DS 65 35-S 7-4405 TLF-DS 65 35-S 7-6230 TLF-DS 65 35-S 7-6461 TLF-DS 65 35-S 7-6882 TLF-DS 65 40-S 7-4248 TLF-DS 65 40-S 7-4250 TLF-DS 65 40-S 7-5480 TLF-DS 65 40-S 7-6228 TLF-DS 65 40-S 7-6554 TLF-DS 65 40-S 7-6740 TLF-DS 65 45-S 7-4582 TLF-DS 65 45-S 7-5484 TLF-DS 65 45-S 7-5808 TLF-DS 65 45-S 7-6668 TLF-DS 65 45-S 7-6742 TLF-DS 65 45-S 7-7299 TLF-DS 65 50-S 7-5488 TLF-DS 65 50-S 7-9264 UDI-DI for associated Model Numbers: TLF-DS 45 25-S 07640185345799 TLF-DS 45 30-S 07640185345805 TLF-DS 45 35-S 07640185345812 TLF-DS 45 40-S 07640185345829 TLF-DS 45 45-S 07640185345836 TLF-DS 55 30-S 07640185345850 TLF-DS 55 35-S 07640185345867 TLF-DS 55 40-S 07640185345874 TLF-DS 55 45-S 07640185345881 TLF-DS 60 35-S 07640305166082

Where was it sold?

U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Spineart

All 13

Other Spinal Implants recalls

All 1,159