CLASS II ONGOING Device recall · Reported 12 Jun 2024 · FDA Enforcement Report

Philips Tempus LS-Manual Defibrillator recalled

Remote Diagnostic Technologies is recalling Philips Tempus LS-Manual Defibrillator, distributed nationwide in the US. The recall was initiated on 9 May 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior
Quantity recalled1745

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2028-2024
FDA event ID
94618
Recalling firm
Remote Diagnostic Technologies, Farnborough, N/A
Classification
Class II
Status
Ongoing
Recall initiated
9 May 2024
FDA classified
6 Jun 2024
Reported
12 Jun 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

9 May 2024Recall initiated by company
6 Jun 2024Classified by FDA+28 days
12 Jun 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

Product as listed by the FDA

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

Where it was distributed

US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.

Nationwide AlabamaArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyMassachusettsMarylandMaineMichiganMinnesotaMissouri

Questions about this recall

Why was it recalled?

Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

Which lots are affected?

00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior

Where was it sold?

US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.