CLASS II ONGOING Device recall · Reported 30 Apr 2025 · FDA Enforcement Report
Tempus Pro Patient Monitor, REFs: 00-1004-R recalled
Remote Diagnostic Technologies is recalling Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset, distributed nationwide in the US. The recall was initiated on 16 Apr 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1666-2025
- FDA event ID
- 96678
- Recalling firm
- Remote Diagnostic Technologies, Farnborough, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 16 Apr 2025
- FDA classified
- 24 Apr 2025
- Reported
- 30 Apr 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 16 Apr 2025 | Recall initiated by company | |
| 24 Apr 2025 | Classified by FDA | +8 days |
| 30 Apr 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with headset dock), 00-1024-R (Tempus Pro printer), 00-1026-R (Tempus Pro printer, IBP), 00-1007 [(Tempus Pro, BT Dock (IB Only)], 00-1004 00-1004 [Tempus Pro (IB Only)]
Where it was distributed
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, KY, MY, NL, NO, OM, PL, PT, RO, SA, SE, SG, SI, SL, TH, ZA.
Nationwide AlabamaArizonaCaliforniaColoradoDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyMassachusettsMarylandMaineMichiganMinnesotaMissouri
Questions about this recall
Why was it recalled?
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
Which lots are affected?
All Lots/ UDI-DI(s): 00-1004-R - 05060472440020 & 05060472442901; 00-1007-R - 05060472440013 & 05060472442918; 00-1024-R - 05060472441027 & 05060472442925; 00-1026-R - 05060472441058 & 05060472442932; 00-1004 - 05060472440624; 00-1007 - 05060472440655
Where was it sold?
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, KY, MY, NL, NO, OM, PL, PT, RO, SA, SE, SG, SI, SL, TH, ZA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Remote Diagnostic Technologies
All 12| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Tempus Pro Patient Monitor recalled | Remote Diagnostic Technologies | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Inseego USB8 Dongle Kit recalled by Remote Diagnostic Technologies | Remote Diagnostic Technologies | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Philips Tempus Pro Patient Monitor recalled | Remote Diagnostic Technologies | Labeling Errors | Nationwide |
| Device | CLASS III | Tempus LS-Manual Defibrillator recalled | Remote Diagnostic Technologies | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Tempus LS-Manual Defibrillator recalled | Remote Diagnostic Technologies | Lead & Heavy Metals | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |