CLASS II ONGOING Device recall · Reported 14 May 2025 · FDA Enforcement Report
Philips Tempus Pro Patient Monitor recalled
Remote Diagnostic Technologies is recalling Philips Tempus Pro Patient Monitor, distributed nationwide in the US. The recall was initiated on 15 Apr 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1729-2025
- FDA event ID
- 96719
- Recalling firm
- Remote Diagnostic Technologies, Farnborough, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 15 Apr 2025
- FDA classified
- 7 May 2025
- Reported
- 14 May 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Labeling Errors
- Device type
- Patient Monitors
Timeline
| 15 Apr 2025 | Recall initiated by company | |
| 7 May 2025 | Classified by FDA | +22 days |
| 14 May 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
Product as listed by the FDA
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Where it was distributed
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.
Nationwide AlabamaArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyMassachusettsMarylandMaineMichiganMinnesota
Questions about this recall
Is Philips Tempus Pro Patient Monitor recalled?
Yes. Remote Diagnostic Technologies recalled Philips Tempus Pro Patient Monitor on 15 Apr 2025. The recall is Class II and its status is Ongoing. Recall number Z-1729-2025.
Why was it recalled?
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
Which lots are affected?
Software versions prior to vx.40. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655
Where was it sold?
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Remote Diagnostic Technologies
All 12| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Tempus Pro Patient Monitor recalled | Remote Diagnostic Technologies | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Inseego USB8 Dongle Kit recalled by Remote Diagnostic Technologies | Remote Diagnostic Technologies | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Tempus Pro Patient Monitor, REFs: 00-1004-R recalled | Remote Diagnostic Technologies | Device Malfunction | Nationwide |
| Device | CLASS III | Tempus LS-Manual Defibrillator recalled | Remote Diagnostic Technologies | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Tempus LS-Manual Defibrillator recalled | Remote Diagnostic Technologies | Lead & Heavy Metals | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |