CLASS II ONGOING Device recall · Reported 4 Nov 2026 · FDA Enforcement Report

Precision Pointer Electromagnetic (8000-050-001) recalled

Stryker Leibinger GmbH & Co. KG is recalling Precision Pointer Electromagnetic (8000-050-001), distributed nationwide in the US. The recall was initiated on 24 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot 7000413361 UDI-DI 07613327413205
Quantity recalled513

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0314-2027
FDA event ID
99855
Recalling firm
Stryker Leibinger GmbH & Co. KG, Freiburg Im Breisgau
Classification
Class II
Status
Ongoing
Recall initiated
24 Sep 2026
FDA classified
29 Oct 2026
Reported
4 Nov 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

24 Sep 2026Recall initiated by company
29 Oct 2026Classified by FDA+35 days
4 Nov 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

Product as listed by the FDA

Precision Pointer Electromagnetic (8000-050-001)

Where it was distributed

Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.

Nationwide AlabamaArizonaCaliforniaColoradoConnecticutDelawareFloridaHawaiiIllinoisIndianaKentuckyLouisianaMassachusettsMichiganMissouriMontanaNew HampshireNew YorkOhioPennsylvania

Questions about this recall

Why was it recalled?

Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

Which lots are affected?

Lot 7000413361 UDI-DI 07613327413205

Where was it sold?

Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 10 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.