CLASS II Device recall · Reported 12 Jan 2022 · FDA Enforcement Report
QIAstat-Dx Respiratory Sars-CoV-2 Panel recalled by Qiagen Sciences
Qiagen Sciences is recalling QIAstat-Dx Respiratory Sars-CoV-2 Panel (US IVD under FDA EUA 200075), distributed in 8 states. The recall was initiated on 8 Nov 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0455-2022
- FDA event ID
- 89130
- Recalling firm
- Qiagen Sciences, Germantown, MD
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 8 Nov 2021
- FDA classified
- 5 Jan 2022
- Reported
- 12 Jan 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- CPAP & Ventilators
Timeline
| 8 Nov 2021 | Recall initiated by company | |
| 5 Jan 2022 | Classified by FDA | +58 days |
| 12 Jan 2022 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
Product as listed by the FDA
QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223
Where it was distributed
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
AlabamaDistrict of ColumbiaKentuckyMarylandMaineNew JerseyPennsylvaniaTexas
Questions about this recall
Why was it recalled?
The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
Which lots are affected?
LOT 210209 UDI (01)14053228038846(17)211127(10)210209(24)691223 Serial Numbers: 412092551 412092559 412092552 412092560 412092553 412092561 412092554 412092563 412092555 412092564 412092556 412092565 412092557 412092566 412092558 412092567
Where was it sold?
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Qiagen Sciences
All 23| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | QIAstat-Dx Respiratory Panel Plus Reference Number recalled | Qiagen Sciences | Device Malfunction | 12 states |
| Device | CLASS II | QIAcube Connect MDx - IVD Instrument designed to perform recalled | Qiagen Sciences | Potency & Assay | Nationwide |
| Device | CLASS II | EZ2 Connect MDx-IVD Designed to perform automated isolation recalled | Qiagen Sciences | Temperature Abuse | Nationwide |
| Device | CLASS II | QIAstat-Dx Respiratory Sars-CoV-2 Panel multiplexed nucleic recalled | Qiagen Sciences | Other | Nationwide |
| Device | CLASS II | QIAcube Connect MDx recalled by Qiagen Sciences | Qiagen Sciences | Other | 26 states |
Other CPAP & Ventilators recalls
All 568| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Vocsn+Pro recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | VC+Pro recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | V*Home recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | V+Pro recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | VC Model Number: Prt-01098-000, Package PN:Prt-01125-000. used recalled | Ventec Life Systems | Packaging Defects | Nationwide |