CLASS II Device recall · Reported 30 Oct 2024 · FDA Enforcement Report
Quantum Perfusion Systems for CPB recalled by Spectrum Medical
Spectrum Medical is recalling Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood, distributed nationwide in the US. The recall was initiated on 13 Mar 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0224-2025
- FDA event ID
- 95428
- Recalling firm
- Spectrum Medical, Gloucester, N/A
- Classification
- Class II
- Status
- Completed
- Recall initiated
- 13 Mar 2023
- FDA classified
- 24 Oct 2024
- Reported
- 30 Oct 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Visit
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 13 Mar 2023 | Recall initiated by company | |
| 24 Oct 2024 | Classified by FDA | +591 days |
| 30 Oct 2024 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
Product as listed by the FDA
Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00
Where it was distributed
US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.
Nationwide AlabamaCaliforniaFloridaKentuckyMassachusettsMaineMinnesotaMissouriNorth CarolinaNew JerseyNew YorkOhioOklahomaPennsylvaniaTennesseeTexasUtahWest Virginia
Questions about this recall
Why was it recalled?
The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
Which lots are affected?
UDI/DI 05060434420770: Serial Numbers: a) AU6004040, AU6004041, AU6004042, AU6004043, AU6004044, AU6004046, AU6004050, AU6004051, AU6004052, AU6004053, AU6004054, AU6004056, AU6004057, AU6004058, AU6004060, AU6004061, AU6004062, AU6004063, AU6004064, AU6004065, AU6004066, AU6004069, AU6004070, AU6004071, AU6004072, AU6004073, AU6004074, AU6004075, AU6004136, AU6004137, AU6004138, AU6004139, AU6004140, AU6004141, AU6004142, AU6004143, AU6004144, AU6004145, AU6004146, AU6004147, AU6004148, AU6004159, AU6004160, AU6004161, AU6004162, AU6004167, AU6004168, AU6004169, AU6004170, AU6004171, AU6004172, AU6004173, AU6004174, AU6004175, AU6004176, AU6004177, AU6004178, AU6004179, AU6004180, AU6004181, AU6004182, AU6004183, AU6004184, AU6004185, AU6004186, AU6004187, AU6004188, AU6004195, AU6004197, AU6004198, AU6004200, AU6004201, AU6004202, AU6004203, AU6004204, AU6004205, AU6004207, AU6004332, AU6004333, AU6004348, AU6004349, AU6004350, AU6004352, AU6004353, AU6004354, AU6004532, AU6004577, AU6004578, AU6004579, AU6004580, AU6004587, AU6004588, AU6004590, AU6004591, AU6004603, AU6004604, AU6004605, AU6004606, AU6004607, AU6004609, AU6004610, AU6004613, AU6004731, AU6004771, AU6004773, AU6004776, AU6004777, AU6004778, AU6004779, AU6004781, AU6004782, AU6004919, AU6004920, AU6004921, AU6004922, AU6004923, AU6004924, AU6004925, AU6004926, AU6004927, AU6004929, AU6004930, AU6004955, AU6004957, AU6004958, AU6004960, AU6004963, AU6004966, AU6004968, AU6004970, AU6004971, AU6004972, AU6005010, AU6005021, AU6005022, AU6005023, AU6005024, AU6005025, AU6005026, AU6005027, AU6005028, AU6005029, AU6005030, AU6005037, AU6005038, AU6005039, AU6005040, AU6005041, AU6005042, AU6005057, AU6005065, AU6005109, AU6005110, AU6005118, AU6005124, AU6005125, AU6005355, AU6006351; Serial Numbers: b) AU8001273, AU8001275, AU8001346, AU8001347, AU8001348, AU8001359, AU8001360, AU8001366, AU8001403, AU8001406, AU8001407, AU8001446, AU8001447, AU8001449
Where was it sold?
US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Spectrum Medical
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Quantum PureFlow Standard Heat Exchanger High Flow 3/8 recalled | Spectrum Medical | Packaging Defects | Nationwide |
| Device | CLASS II | Quantum PureFlow Standard Heat Exchanger High Flow 3/8 recalled | Spectrum Medical | Packaging Defects | Nationwide |
| Device | CLASS II | Quantum Informatics - Vision / Live Vue Monitor recalled | Spectrum Medical | Device Malfunction | 38 states |
| Device | CLASS II | Quantum Informatics - Viper Monitor, Physiological, Patient recalled | Spectrum Medical | Device Malfunction | 38 states |
| Device | CLASS II | Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type recalled | Spectrum Medical | Packaging Defects | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |