CLASS II ONGOING Device recall · Reported 12 Feb 2025 · FDA Enforcement Report
The Rapid fFN Specimen Collection Tube (PN recalled by Hologic
Hologic is recalling The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN, distributed nationwide in the US. The recall was initiated on 8 Jan 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1095-2025
- FDA event ID
- 96116
- Recalling firm
- Hologic, San Diego, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 8 Jan 2025
- FDA classified
- 4 Feb 2025
- Reported
- 12 Feb 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Test Kits & Diagnostics
Timeline
| 8 Jan 2025 | Recall initiated by company | |
| 4 Feb 2025 | Classified by FDA | +27 days |
| 12 Feb 2025 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN¿ for the TLiIQ¿ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.
Where it was distributed
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Guam and the countries of Canada and Japan.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland
Questions about this recall
Is The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN recalled?
Yes. Hologic recalled The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN on 8 Jan 2025. The recall is Class II and its status is Ongoing. Recall number Z-1095-2025.
Why was it recalled?
Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN¿ for the TLiIQ¿ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
Which lots are affected?
Lot # 901164, 901165, 904850/ UDI: 25420045514765
Where was it sold?
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Guam and the countries of Canada and Japan.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hologic
All 48| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | 3dimensions Mammography System recalled by Hologic | Hologic | Other | Nationwide |
| Device | CLASS II | Selenia Dimensions Mammography Systems recalled by Hologic | Hologic | Other | Nationwide |
| Device | CLASS II | Genius" Review Station Display (Barco Monitor) recalled by Hologic | Hologic | Device Malfunction | Nationwide |
| Device | CLASS II | Aptima HPV Assay recalled by Hologic | Hologic | Potency & Assay | Nationwide |
| Device | CLASS II | Brevera Breast Biopsy System Disposable Needle recalled by Hologic | Hologic | Particulate Matter | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |