CLASS II Device recall · Reported 18 Nov 2015 · FDA Enforcement Report

Samsung GM60A-40S Intended for use recalled by NeuroLogica

NeuroLogica is recalling Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy, distributed in Massachusetts, Missouri, North Carolina, Ohio. The recall was initiated on 14 Sep 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F
Quantity recalled7 devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0263-2016
FDA event ID
72171
Recalling firm
NeuroLogica, Danvers, MA
Classification
Class II
Status
Terminated
Recall initiated
14 Sep 2015
FDA classified
10 Nov 2015
Reported
18 Nov 2015
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

14 Sep 2015Recall initiated by company
10 Nov 2015Classified by FDA+57 days
18 Nov 2015Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Where it was distributed

Distributed to the states of NC, OH, MA, and MO.

MassachusettsMissouriNorth CarolinaOhio

Questions about this recall

Why was it recalled?

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Which lots are affected?

Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F

Where was it sold?

Distributed to the states of NC, OH, MA, and MO.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from NeuroLogica

All 13

Other Imaging Equipment recalls

All 2,347