CLASS I ONGOING Device recall · Reported 3 Jul 2024 · FDA Enforcement Report
Sense XL Torso Coil 3.0t recalled by Philips North America
Philips North America is recalling Sense XL Torso Coil 3.0t, distributed nationwide in the US. The recall was initiated on 5 Jun 2024 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2164-2024
- FDA event ID
- 94786
- Recalling firm
- Philips North America, Cambridge, MA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 5 Jun 2024
- FDA classified
- 27 Jun 2024
- Reported
- 3 Jul 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Imaging Equipment
Timeline
| 5 Jun 2024 | Recall initiated by company | |
| 27 Jun 2024 | Classified by FDA | +22 days |
| 3 Jul 2024 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for coils to heat up and harm patients (burn).
Product as listed by the FDA
SENSE XL TORSO COIL 3.0T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567394941, 453567394942, 453567394943, 453567394945, 459801593182. Coils may be included in kits with Mode Numbers: 989603050641, 989603050642.
Where it was distributed
Worldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Papua New Guinea, Peru, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, T¿rkiye, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
Questions about this recall
Why was it recalled?
Potential for coils to heat up and harm patients (burn).
Which lots are affected?
UDI-DI: N/A. 1. Model Number 453567394941. Serial Number: 9. 2. Model Number 453567394942. Serial Numbers: 40000, 100, 102, 104, 144, 192, 84, 98, 42. 3. Model Number 453567394943. Serial Numbers: 301, 228, 761, 87, 880, 814, 0174R3, 541, @001@0179R3, 0141R3, 0172R3, 0189R3, 0260R3, 0265R3, 0285R3, 0358R3, 116R3, 121, 127, 149, 152, 158, 165, 176, 180, 213R3, 216, 220, 224, 235, 237, 263, 265, 267, 273, 274, 286, 294, 298, 302, 303, 310R3, 312, 313, 316, 328R3, 336, 346, 359, 360, 374, 377, 382, 389, 400, 406, 421, 423, 424, 43, 431, 45, 454, 455, 460, 464, 466, 474, 477, 481, 488, 496, 505, 511, 513, 518, 523, 543, 549, 564, 576, 582, 592, 597, 602, 603, 61, 610, 611, 617R3, 627, 632, 633, 635, 642, 657, 658, 666r3, 669, 670, 672, 685, 686, 694, 708, 710, 716, 720, 726, 729, 732, 736, 745, 757, 761, 763, 780, 782, 783, 788, 793, 800, 828, 84, 861, 87, 880, 887, 89, 907, 908, 913, 924, 926, 932, 939, 947, 956, 961, 967. 4. Model Number 453567394945. Serial Numbers: 104, 107, 108, 117, 129, 145, 165, 166, 178, 180, 198, 201, 211, 22, 222, 225, 229, 242, 268, 31, 336, 35, 42, 64, 69, 87, 916, 94. 5. Model Number 459801593182. Serial Numbers: 116, 141, 165, 211, 100, 103, 104, 110, 112, 123, 124, 127, 128, 13, 131, 133, 134, 136, 142, 143, 144, 145, 146, 148, 15, 150, 156, 158, 164, 167, 168, 172, 173, 174, 179, 183, 186, 192, 193, 194, 195, 198, 199, 20, 204, 208, 21, 217, 221, 225, 23, 230, 232, 240, 241, 246, 249, 251, 253, 254, 26, 262, 264, 265, 266, 268, 275, 278, 279, 28, 288, 294, 297, 298, 303, 306, 311, 312, 313, 315, 321, 324, 326, 330, 339, 34, 340, 344, 345, 346, 349, 350, 352, 354, 355, 357, 361, 362, 365, 368, 370, 39, 40, 42586, 44, 52, 53, 57, 6, 60, 62, 63, 64, 66, 705, 71, 72, 79, 80, 88, 932, 94, 97. 6. Model Number 989603050641. Serial Numbers: 152, 169, 179, 88, 83, 109, 105, 130, 176. 7. Model Number 989603050642. Serial Numbers: 529, 509, 217, 189, 127, 311, 85, 531, 534, 546, 561, 347, 609, 296, 142, 110, 391, 137, 20, 28, 160, 378, 370, 172, 202, 76, 272, 326, 404, 167, 94, 446, 560, 521, 497, 353, 515, 386, 369, 253, 571, 220, 333, 362, 265, 426, 442, 469, 74, 31, 284, 533, 147, 248, 259, 69, 330, 270, 525, 611, 102, 309, 461, 52, 191, 203, 108, 8, 100, 524, 381, 503, 389, 283, 360, 53, 348, 448, 490, 556, 22, 380, 247, 59, 554, 589, 580, 403, 507, 226, 452, 183, 357, 224, 472, 39, 221, 435, 16, 480, 266, 255, 445, 588, 1, 128, 466, 32, 124, 133, 276, 329, 431, 443, 598, 227, 584, 14, 449, 295, 513, 168, 535, 40, 17, 48, 104, 51, 177, 64, 287, 9, 394, 174, 494, 201, 50, 482, 590, 350, 438, 487, 365, 297, 368, 57, 196, 432, 551, 310, 12, 242, 250, 249, 240, 366, 382, 18, 390, 320, 129, 604, 314, 545, 437, 58, 583, 175, 367, 274, 510, 512, 573, 131, 505, 44, 599, 47, 595, 398, 206, 593, 564, 495, 619, 54, 30, 122, 552, 312, 161, 254, 275, 282, 387, 123, 335, 364, 570, 607, 166, 91, 384, 379, 436, 508, 193, 474, 407, 575, 577, 608, 601, 356, 188, 232, 324, 430, 422, 455, 155, 195, 321, 300, 361, 194, 208, 338, 418, 214, 99, 243, 492, 317, 581, 103, 237, 340, 401.
Where was it sold?
Worldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Papua New Guinea, Peru, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, T¿rkiye, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 94786This product is part of a recall event; the main page for the event is Sense XL Torso Coil 1.5t recalled by Philips North America.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | Sense XL Torso Coil 1.5t recalled by Philips North America | Philips North America | Other | Nationwide | |
| CLASS I | Sense XL Torso Coil 1.5T Mk2 recalled by Philips North America | Philips North America | Other | Nationwide |
More recalls from Philips North America
All 531| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel recalled | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Cardiac Workstation recalled by Philips North America | Philips North America | Labeling Errors | Nationwide |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederl | Potency & Assay |