CLASS II ONGOING Device recall · Reported 18 Jun 2025 · FDA Enforcement Report

Solana¿ Bordetella Complete Assay Solana Bordetella kit recalled

Ortho-Clinical Diagnostics is recalling Solana¿ Bordetella Complete Assay Solana Bordetella kit. The recall was initiated on 24 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: Catalog number: M308 UDI-DI: 30014613336167; Lots: 245251 - Exp. 2025-Oct-25 246941 - Exp. 2025-Dec-03 250163 - Exp. 2026-Feb-06 259686 - Exp. 2026-Sep-03 261072 - Exp. 2026-Oct-02 262113 - Exp. 2026-Oct-23 263331 - Exp. 2026-Nov-22 264065 - Exp. 2026-Dec-11 265007 - Exp. 2027-Jan-03 (corrected exp date from 2026-Dec-31) 266928 - Exp. 2026-Dec-31 Additional Lots: 265007-1 - Exp. 2026-Dec-31 268479 and above - Various Expiration dates *Until further notice, all future lots, both supplied by QuidelOrtho or supplied by its distributors, will be affected.
Quantity recalled952 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1948-2025
FDA event ID
96870
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
24 Mar 2025
FDA classified
11 Jun 2025
Reported
18 Jun 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

24 Mar 2025Recall initiated by company
11 Jun 2025Classified by FDA+79 days
18 Jun 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component

Where it was distributed

Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho sent the notification to their consignees because QuidelOrtho drop-shipped this product to their 31 consignees (Included in the Medical Facility total below).

Questions about this recall

Why was it recalled?

It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.

Which lots are affected?

Lot Code: Catalog number: M308 UDI-DI: 30014613336167; Lots: 245251 - Exp. 2025-Oct-25 246941 - Exp. 2025-Dec-03 250163 - Exp. 2026-Feb-06 259686 - Exp. 2026-Sep-03 261072 - Exp. 2026-Oct-02 262113 - Exp. 2026-Oct-23 263331 - Exp. 2026-Nov-22 264065 - Exp. 2026-Dec-11 265007 - Exp. 2027-Jan-03 (corrected exp date from 2026-Dec-31) 266928 - Exp. 2026-Dec-31 Additional Lots: 265007-1 - Exp. 2026-Dec-31 268479 and above - Various Expiration dates *Until further notice, all future lots, both supplied by QuidelOrtho or supplied by its distributors, will be affected.

Where was it sold?

Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho sent the notification to their consignees because QuidelOrtho drop-shipped this product to their 31 consignees (Included in the Medical Facility total below).

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656