CLASS II Device recall · Reported 3 Aug 2016 · FDA Enforcement Report

Stryker Orthopaedics Trident Constrained Liner recalled

Stryker Howmedica Osteonics is recalling Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, and Non Sterile, distributed nationwide in the US. The recall was initiated on 3 Jun 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Nos. 2199-2022, 2199-2028 and 2199-2032 All lots manufactured from May 30, 2012
Quantity recalled1,864 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2281-2016
FDA event ID
74500
Recalling firm
Stryker Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Terminated
Recall initiated
3 Jun 2016
FDA classified
25 Jul 2016
Reported
3 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

3 Jun 2016Recall initiated by company
25 Jul 2016Classified by FDA+52 days
3 Aug 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.

Where it was distributed

Worldwide distribution. US nationwide (AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OH, PA, TX, WA, WI, WV), Sweden, South Africa, Chile, Australia, New Zealand, Canada, United Kingdom, Switzerland, Colombia, and Poland.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMississippiNew Jersey

Questions about this recall

Why was it recalled?

Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.

Which lots are affected?

Catalog Nos. 2199-2022, 2199-2028 and 2199-2032 All lots manufactured from May 30, 2012

Where was it sold?

Worldwide distribution. US nationwide (AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OH, PA, TX, WA, WI, WV), Sweden, South Africa, Chile, Australia, New Zealand, Canada, United Kingdom, Switzerland, Colombia, and Poland.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Howmedica Osteonics