CLASS II Device recall · Reported 1 Nov 2017 · FDA Enforcement Report

Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain recalled

Sysmex America is recalling Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood, distributed nationwide in the US. The recall was initiated on 8 Sep 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Numbers: A6051 A6052 A6053 A6054 A6055 A6056 A6057 A6058 A6059 A6060 A6061 A6062 A6063 A6064 A6065 A6066 A6067 A6068 A6069 A6070 A6071 A6072 A6073 A6074 A6075 A6076 A6077 A6078 A6079 A6080 A6081 A6082 A6083 A6084 A6085 A6086 A6087 A6088 A6089 A6090
Quantity recalled23,239 units (9353 US)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0045-2018
FDA event ID
78134
Recalling firm
Sysmex America, Lincolnshire, IL
Classification
Class II
Status
Completed
Recall initiated
8 Sep 2017
FDA classified
24 Oct 2017
Reported
1 Nov 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

8 Sep 2017Recall initiated by company
24 Oct 2017Classified by FDA+46 days
1 Nov 2017Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.

Product as listed by the FDA

Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers

Where it was distributed

Worldwide Distribution - US (nationwide including Puerto Rico) Internationally to New Zealand, Brazil, India, Philippines, Nicaragua, Argentina, Australia, Chile, Viet Nam, Colombia, Pakistan, Malaysia, Dubai, Germany, Hong Kong, Bangladesh, Uruguay Thailand, China, Saudi Arabia, Taiwan, Honduras, Ecuador, Kuwait, South Korea, and Costa Rica

Nationwide Puerto Rico

Questions about this recall

Is Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood recalled?

Yes. Sysmex America recalled Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood on 8 Sep 2017. The recall is Class II and its status is Completed. Recall number Z-0045-2018.

Why was it recalled?

There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.

Which lots are affected?

Lot Numbers: A6051 A6052 A6053 A6054 A6055 A6056 A6057 A6058 A6059 A6060 A6061 A6062 A6063 A6064 A6065 A6066 A6067 A6068 A6069 A6070 A6071 A6072 A6073 A6074 A6075 A6076 A6077 A6078 A6079 A6080 A6081 A6082 A6083 A6084 A6085 A6086 A6087 A6088 A6089 A6090

Where was it sold?

Worldwide Distribution - US (nationwide including Puerto Rico) Internationally to New Zealand, Brazil, India, Philippines, Nicaragua, Argentina, Australia, Chile, Viet Nam, Colombia, Pakistan, Malaysia, Dubai, Germany, Hong Kong, Bangladesh, Uruguay Thailand, China, Saudi Arabia, Taiwan, Honduras, Ecuador, Kuwait, South Korea, and Costa Rica

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.