CLASS II Device recall · Reported 3 Oct 2018 · FDA Enforcement Report

Terumo 220/240V AC, Advanced Perfusion System Platform recalled

Terumo Cardiovascular Systems is recalling Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary, distributed nationwide in the US. The recall was initiated on 28 Aug 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 1522 1523 1524 1525 1528 1530 1526 1527 1532 1544 Spare Parts: 870509 ASSEMBLY POWER MANAGER POD APS S/N: 03097 03099 03100 03101 03113 03114 03115 03143 03155 03167 03264
Quantity recalled10 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3243-2018
FDA event ID
80957
Recalling firm
Terumo Cardiovascular Systems, Ann Arbor, MI
Classification
Class II
Status
Terminated
Recall initiated
28 Aug 2018
FDA classified
25 Sep 2018
Reported
3 Oct 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

28 Aug 2018Recall initiated by company
25 Sep 2018Classified by FDA+28 days
3 Oct 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system

Product as listed by the FDA

Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764

Where it was distributed

Nationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam

Nationwide

Questions about this recall

Why was it recalled?

Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system

Which lots are affected?

Serial Numbers: 1522 1523 1524 1525 1528 1530 1526 1527 1532 1544 Spare Parts: 870509 ASSEMBLY POWER MANAGER POD APS S/N: 03097 03099 03100 03101 03113 03114 03115 03143 03155 03167 03264

Where was it sold?

Nationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 80957
ReportedClassRecallCompanyReasonWhere
CLASS IITerumo 100/120V AC, Advanced Perfusion System Platform recalledTerumo Cardiovascular SystemsOtherNationwide

More recalls from Terumo Cardiovascular Systems

All 222

Other Other Medical Devices recalls

All 13,884