CLASS II Device recall · Reported 23 Sep 2020 · FDA Enforcement Report

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3//32 recalled

Terumo Cardiovascular Systems is recalling Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3//32, distributed nationwide in the US. The recall was initiated on 25 Aug 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 00886799000038; Serial Numbers: 20288 20289 20290 20291 20292 20293 20294 20295 20296 20299 20300 20301 20302 20306 20307 20308 20309 20310 20312 20317 20318 20320 20321 20322 20323 20324 20326 20328 20329 20331 20332 20334 20338 20339 20341 20342 20343 20344 20345 20346 20347 20348 20349 20350 20351 20352 20358 20365 20368 20373 20374 20375 20381 20382 20383 20384 20385 20387 20388 20392 20393 20395 20396 20397 20398 20399 20401 20402 20403 20404 20405 20408 20410 20411 20412 20555 20556 20557 20558 20559 20560 20561 20562 20563 20564 20565 20566 20567 20568 20569 20570 20571 20572 20573 20574 20575 20576 20577 20578 20579
Quantity recalled100

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2959-2020
FDA event ID
86338
Recalling firm
Terumo Cardiovascular Systems, Ann Arbor, MI
Classification
Class II
Status
Terminated
Recall initiated
25 Aug 2020
FDA classified
11 Sep 2020
Reported
23 Sep 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Aug 2020Recall initiated by company
11 Sep 2020Classified by FDA+17 days
23 Sep 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Where it was distributed

Worldwide distribution - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Which lots are affected?

UDI: 00886799000038; Serial Numbers: 20288 20289 20290 20291 20292 20293 20294 20295 20296 20299 20300 20301 20302 20306 20307 20308 20309 20310 20312 20317 20318 20320 20321 20322 20323 20324 20326 20328 20329 20331 20332 20334 20338 20339 20341 20342 20343 20344 20345 20346 20347 20348 20349 20350 20351 20352 20358 20365 20368 20373 20374 20375 20381 20382 20383 20384 20385 20387 20388 20392 20393 20395 20396 20397 20398 20399 20401 20402 20403 20404 20405 20408 20410 20411 20412 20555 20556 20557 20558 20559 20560 20561 20562 20563 20564 20565 20566 20567 20568 20569 20570 20571 20572 20573 20574 20575 20576 20577 20578 20579

Where was it sold?

Worldwide distribution - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 86338

This product is part of a recall event; the main page for the event is Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1//32 recalled.

More recalls from Terumo Cardiovascular Systems

All 222

Other Other Medical Devices recalls

All 13,884