CLASS II Device recall · Reported 23 Sep 2020 · FDA Enforcement Report

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor recalled

Terumo Cardiovascular Systems is recalling Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1//16 UAS BLK, distributed nationwide in the US. The recall was initiated on 25 Aug 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 00886799000045; Serial Numbers: 20230 20263 20265
Quantity recalled3

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2960-2020
FDA event ID
86338
Recalling firm
Terumo Cardiovascular Systems, Ann Arbor, MI
Classification
Class II
Status
Terminated
Recall initiated
25 Aug 2020
FDA classified
11 Sep 2020
Reported
23 Sep 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Aug 2020Recall initiated by company
11 Sep 2020Classified by FDA+17 days
23 Sep 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x1/16 UAS BLK, Catalog Number 5785 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Where it was distributed

Worldwide distribution - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Which lots are affected?

UDI: 00886799000045; Serial Numbers: 20230 20263 20265

Where was it sold?

Worldwide distribution - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 86338

This product is part of a recall event; the main page for the event is Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1//32 recalled.

More recalls from Terumo Cardiovascular Systems

All 222

Other Other Medical Devices recalls

All 13,884