CLASS II Device recall · Reported 12 Mar 2014 · FDA Enforcement Report
TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System recalled
Toshiba American Medical Systems is recalling TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System, distributed nationwide in the US. The recall was initiated on 1 Oct 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1113-2014
- FDA event ID
- 66527
- Recalling firm
- Toshiba American Medical Systems, Tustin, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 1 Oct 2013
- FDA classified
- 3 Mar 2014
- Reported
- 12 Mar 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 1 Oct 2013 | Recall initiated by company | |
| 3 Mar 2014 | Classified by FDA | +153 days |
| 12 Mar 2014 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Where it was distributed
Nationwide Distribution.
Questions about this recall
Why was it recalled?
Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
Which lots are affected?
SERIAL NUMBER JDA1062176, JDA1062177, JDA1072183, JDA1072184, JDA1072186, JDA1072187, JDA1082188, JDA1082189, JDA1082190, JDA10X2191, JDA10X2192, JDA10X2193, JDA10X2194, JDA10X2195, JDA10X2196, JDA10X2197, JDA10X2198, JDA10Y2199, JDA10Y2200, JDA10Y2201, JDA10Y2202, JDA10Y2203, JDA10Y2204, JDA10Y2205, JDA10Z2206, JDA10Z2207, JDA10Z2209, JDA10Z2210, JDA10Z2211, JDA10Z2212, JDA10Z2213, JDA10Z2214, JDA10Z2215, JDA1112216, JDA1112217, JDA1112218, JDA1112219, JDA1112221, JDA1112222, JDA1112224, JDA1112225, JDA1112226, JDA1112227, JDA1122231, JDA1122234, JDA1142236, JDA1162237, JDA1162238, JDA1162239, JDA1182240, JDA1182241, JDA1182242, JDA1182243, JDA1182244, JDA11X2245, JDA11X2246, JDA11X2247, JDA11Z2249, JDA11Z2250, JDA11Z2252, JDA11Z2253, JDA11Z2254, JDA11Z2255, JDA1222257, JDA1222258, JDA1222259, JDA1222260, JDA1222261, JDA1242264, JDA1252265, JDA1252266, JDA1252267, JDA12Z2271, JDA1312272, JGA1062178, JGA1062181, JGA1062182, JGA1112220, JGA1112228, JGA11Z2256, JGA1222262, JGA12X2269, JLA1062179, JLA1062180, JLA1072185, JLA10Z2208, JLA1122232, JLA1122233, JLA1132235, JLA11Y2248, JLA11Z2251, JLA1232263, JLA1262268, JLA12Y2270, KDA1062080, KDA1072081, KDA1072082, KDA1072083, KDA1072085, KDA1072086, KDA1072088, KDA1082090, KDA1082091, KDA1082092, KDA1092093, KDA10X2094, KDA10X2095, KDA10X2096, KDA10X2097, KDA10X2098, KDA10X2099, KDA10Y2100, KDA10Y2101, KDA10Y2102, KDA10Z2104, KDA1112107, KDA1122110, KDA1122111, KDA1132113, KDA1142114, KDA1152115, KDA1152116, KDA1152117, KDA1152118, KDA1162119, KDA1162120, KDA1162121, KDA1162122, KDA1162123, KDA1162124, KDA1172125, KDA1182126, KDA1182127, KDA1182128, KDA1182129, KDA1192130, KDA1192131, KDA11X2132, KDA11X2133, KDA11Y2134, KDA11Z2136, KDA11Z2137, KDA11Z2138, KDA11Z2139, KDA11Z2140, KDA1222141, KDA1222142, KDA1222143, KDA1232144, KDA1242145, KDA1242146, KDA1252147, KDA1272148, KDA1282149, KDA1292151, KDA12X2152, KDA12Y2153, KDA12Z2154, KDA1322155, KGA1082089, KGA10Z2103, KGA10Z2105, KGA1112106, KGA1112108, KGA1122109, KGA1122112, KGA11Z2135, KLA1072084, KLA1072087, KLA1292150, RDA1292019, RDA12X2023, RDA12X2024, RDB12Y2029, RDB12Z2032, RDB1312037, RDB1312038, RGA12X2025, RGB12Y2028, RGB1312034, 2DA0792002, 2DA0792004, 2DA07X2005, 2DA07X2008, 2DA07Y2010, 2DA0842023, 2DA0842026, 2DA0862034, 2DA0862036, 2DA0872038, 2DA0882048, 2DA0882049, 2DA0882051, 2DA0882052, 2DA0882053, 2DA08X2061, 2DA08X2067, 2DA08X2070, 2DA08X2071, 2DA08X2072, 2DA08Y2077, 2DA08Y2080, 2DA08Z2092, 2DA0952115, 2DA0962116, 2DA0972123, 2DA0982131, 2DA0982133, 2DA09X2139, 2DA09X2143, 2DA09Y2148, 2DA09Y2151, 2DA09Z2158, 2DA09Z2161, 2DA09Z2162, 2DA1012165, 2DA1012167, 2DA1012168, 2DA1012169, 2DA1012170, 2DA1012174, 2DA1062208, 2DA1072212, 2DB1082219, 2DB1082220, 2DB1082222, 2DB10X2241, 2DB10Y2253, 2DB10Y2257, 2DB10Z2265, 2DB1112275, 2DB1112283, 2DB1122284, 2DB1122291, 2DB1122294, 2DC1142308, 2DC1142311, 2DC1152316, 2DC1162324, 2DC1172328, 2DC1172332, 2DC1172333, 2DC11X2352, 2DC11Y2365, 2DC11Y2367, 2DC11Z2372, 2DC11Z2374, 2DC1212377, 2DC1212383, 2DC1232392, 2DC1232393, 2DC1232396, 2DC1232397, 2DC1232398, 2DC1242407, 2DC1242409, 2DC1252410, 2DC1262412, 2DC1272417, 2DC1272421, 2DC1282423, 2DC12X2427, 2DC12Y2430, 2DC12Y2431, 2DC12Z2433, 2DC1312435, 2DC1312436, 2DC1322437, 1DA0962006, 1DA0972007, 1DA0972008, 1DA0992011, 1DA09Y2014, 1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019, 1DB10Y2020, 1DB10Y2021, 1DB10Y2022, 1DB1112025, 1DB1122027, 1DC1142030, 1DC1152031, 1DC1162032, 1DC1162033, 1DC1172035, 1DC11Y2037, 1DC11Y2038, 1DC11Z2040, 1DC1222042, 1DC1232043, 1DC1232044, 1DC1252045, 1DC1252046, 1DC1262048, 1DC1272050, 1DC1282051, 1DC1292052, 1DC12X2053, 1DC12Y2054, 1DC12Y2055, 1DC12Z2056, 1DC1312057, 1DC1312058, 1DC1342059, 1GC1262047, 1GC1272049, 1DA1222096, 1DA1222097, 1DA1232099, 1DA1232100, 1DA1232103, 1DA1252124, 1DA1252131, 2DA1252001, 2DA1262003, 2DA1272004, 2DA1272005, 2DA1272006, 2DA1282007, 2DA1292008, 2DA12X2009, 2DA12X2010, 2DA12X2012, 2DA12X2013, 2DA12Y2014, 2DA12Y2015, 2DA12Y2016, 2DA12Z2017, 2DA12Z2018, 2DA1312019, 2DA1312020, 2DA1312021, 2DA1312022, 2GA1262002, 2GA12X2011, 2DA1072211, 2DA08X2063, 2DC11Y2368.
Where was it sold?
Nationwide Distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Toshiba American Medical Systems
All 57| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Diagnostic Ultrasound System, Aplio I-series recalled | Toshiba American Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Infx-8000V Fluoroscopic X-Ray Systems recalled | Toshiba American Medical Systems | Other | CA |
| Device | CLASS II | Infx-8000H Fluoroscopic X-Ray Systems recalled | Toshiba American Medical Systems | Other | CA |
| Device | CLASS II | Infx-8000F Fluoroscopic X-Ray Systems recalled | Toshiba American Medical Systems | Other | CA |
| Device | CLASS II | Infx-8000C Fluoroscopic X-Ray Systems recalled | Toshiba American Medical Systems | Other | CA |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederl | Potency & Assay |