CLASS II Device recall · Reported 17 Apr 2013 · FDA Enforcement Report

Tsx-ioi: Aquilion 32/64 recalled by Toshiba American Medical

Toshiba American Medical Systems is recalling Tsx-ioi: Aquilion 32/64, distributed nationwide in the US. The recall was initiated on 30 Nov 2011 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: HDA0722595 JDA0972130 JDA0972135 JDA0992147 JDA09Y2158 KDA09X2063 JDA09Y2157 JDA09Y2161 JDA1012171 JDA10X2191 JDA10X2197 JDA10Z2214 JGAl112220 JLA1122232 JLA1132235 JDAl112216 JDAl112219 NLA0972032 JDA1182242 JDA1182241 JDA1182244 JDA1072183 JDA1082189
Quantity recalled23 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1067-2013
FDA event ID
64667
Recalling firm
Toshiba American Medical Systems, Tustin, CA
Classification
Class II
Status
Terminated
Recall initiated
30 Nov 2011
FDA classified
7 Apr 2013
Reported
17 Apr 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Nov 2011Recall initiated by company
7 Apr 2013Classified by FDA+494 days
17 Apr 2013Published in enforcement report+10 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm initiated this recall due to a potential software issue. The DLP (Dose Length Product) value may be displayed incorrectly when the vHP option is used.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

TSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of a human body.

Where it was distributed

Nationwide Distribution including the states of WA, OH, CA, TX, WV, IL, AZ, CA, WI, ME, WI, NC, FL, MT, and HI.

Nationwide ArizonaCaliforniaFloridaHawaiiIllinoisMaineMontanaNorth CarolinaOhioTexasWashingtonWisconsinWest Virginia

Questions about this recall

Why was it recalled?

The firm initiated this recall due to a potential software issue. The DLP (Dose Length Product) value may be displayed incorrectly when the vHP option is used.

Which lots are affected?

Serial Numbers: HDA0722595 JDA0972130 JDA0972135 JDA0992147 JDA09Y2158 KDA09X2063 JDA09Y2157 JDA09Y2161 JDA1012171 JDA10X2191 JDA10X2197 JDA10Z2214 JGAl112220 JLA1122232 JLA1132235 JDAl112216 JDAl112219 NLA0972032 JDA1182242 JDA1182241 JDA1182244 JDA1072183 JDA1082189

Where was it sold?

Nationwide Distribution including the states of WA, OH, CA, TX, WV, IL, AZ, CA, WI, ME, WI, NC, FL, MT, and HI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Toshiba American Medical Systems

All 57

Other Imaging Equipment recalls

All 2,336