CLASS II ONGOING Device recall · Reported 24 Sep 2025 · FDA Enforcement Report

UniCel DxH 900 Coulter Cellular Analysis System recalled

Beckman Coulter is recalling UniCel DxH 900 Coulter Cellular Analysis System, configured with the Led HGB Photometer, distributed nationwide in the US. The recall was initiated on 13 Aug 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 15099590701574/ Serial Numbers: BB33238, BB33239, BB42317, BC05150, BC08198, BC10226, BC16320, BC21392, BC23453, BC25457, BC25465, BC30527, BC32542, BC42721, BC42726, BC49882, BD13219, BD41570, BD46639, BD46643, BD46646, BE01022, BE02044, BE02052, BE04080, BE21222, BE24254, BE24257, BE29334, BE31336, BE33366, BE42455, BE42459, BE46517, BE49539, BE49541, BF10071, BF12126, BF12129, BF18194, BF33377, BF40462, BG04041, BG04042, BG04053, BG05080, BG11182, BG11184, BG11185, BG14195, BG14200, BG14204, BG14208, BG16220, BG16231, BG17238, BG17250, BG17260, BG19283, BG20292, BG20294, BG25395, BG25412, BG25413, BG26435, BG40563, BG45631, BG46635, BH04028, BH04029, BH05051, BH19220, BH24260, BH31328, BD46642, BG22347, BG20307, BD46640, BC32545
Quantity recalledN/A

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2572-2025
FDA event ID
97464
Recalling firm
Beckman Coulter, Brea, CA
Classification
Class II
Status
Ongoing
Recall initiated
13 Aug 2025
FDA classified
17 Sep 2025
Reported
24 Sep 2025
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

13 Aug 2025Recall initiated by company
17 Sep 2025Classified by FDA+35 days
24 Sep 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

Product as listed by the FDA

UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478

Where it was distributed

Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.

Nationwide AlabamaArizonaCaliforniaConnecticutFloridaGeorgiaIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiMontanaNorth Carolina

Questions about this recall

Why was it recalled?

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

Which lots are affected?

UDI: 15099590701574/ Serial Numbers: BB33238, BB33239, BB42317, BC05150, BC08198, BC10226, BC16320, BC21392, BC23453, BC25457, BC25465, BC30527, BC32542, BC42721, BC42726, BC49882, BD13219, BD41570, BD46639, BD46643, BD46646, BE01022, BE02044, BE02052, BE04080, BE21222, BE24254, BE24257, BE29334, BE31336, BE33366, BE42455, BE42459, BE46517, BE49539, BE49541, BF10071, BF12126, BF12129, BF18194, BF33377, BF40462, BG04041, BG04042, BG04053, BG05080, BG11182, BG11184, BG11185, BG14195, BG14200, BG14204, BG14208, BG16220, BG16231, BG17238, BG17250, BG17260, BG19283, BG20292, BG20294, BG25395, BG25412, BG25413, BG26435, BG40563, BG45631, BG46635, BH04028, BH04029, BH05051, BH19220, BH24260, BH31328, BD46642, BG22347, BG20307, BD46640, BC32545

Where was it sold?

Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 97464

This product is part of a recall event; the main page for the event is UniCel DxH 800 Coulter Cellular Analysis System recalled.

More recalls from Beckman Coulter

All 334

Other Other Medical Devices recalls

All 13,857