CLASS II Device recall · Reported 26 Oct 2016 · FDA Enforcement Report

Vitros Chemistry Products Uric Slides recalled

Ortho-Clinical Diagnostics is recalling Vitros Chemistry Products Uric Slides, distributed nationwide in the US. The recall was initiated on 11 Aug 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCurrent/In-date GENs 26, 27, 28, 29, 31
Quantity recalledUS: 18,749 sales units; Foreign: 94,928 sales units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0092-2017
FDA event ID
74950
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Terminated
Recall initiated
11 Aug 2016
FDA classified
17 Oct 2016
Reported
26 Oct 2016
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Potency & Assay

Timeline

11 Aug 2016Recall initiated by company
17 Oct 2016Classified by FDA+67 days
26 Oct 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier No. 10758750000340, Rx ONLY, IVD --- For in vitro diagnostic use only.

Where it was distributed

Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.

Nationwide Puerto Rico

Questions about this recall

Is Vitros Chemistry Products Uric Slides recalled?

Yes. Ortho-Clinical Diagnostics recalled Vitros Chemistry Products Uric Slides on 11 Aug 2016. The recall is Class II and its status is Terminated. Recall number Z-0092-2017.

Why was it recalled?

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.

Which lots are affected?

Current/In-date GENs 26, 27, 28, 29, 31

Where was it sold?

Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656