CLASS III Device recall · Reported 20 Jul 2016 · FDA Enforcement Report

Vitros Immunodiagnostic Products AFP Reagent Pack recalled

Ortho-clinical Diagnostics is recalling Vitros Immunodiagnostic Products AFP Reagent Pack, distributed nationwide in the US. The recall was initiated on 9 May 2016 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesLots 6471 (expiry date 19-Jul-2016), 6480 (expiry date 23-Aug-2016)
Quantity recalledUS: Lot 6471 - 282, Lot 6480 - 344

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2145-2016
FDA event ID
74346
Recalling firm
Ortho-clinical Diagnostics, Bridgend
Classification
Class III
Status
Terminated
Recall initiated
9 May 2016
FDA classified
8 Jul 2016
Reported
20 Jul 2016
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction

Timeline

9 May 2016Recall initiated by company
8 Jul 2016Classified by FDA+60 days
20 Jul 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ortho Clinical Diagnostics (OCD) confirmed that specific lots of VITROS Immunodiagnostic Products AFP Reagent Pack product exhibit an increase in the reaction signal over time potentially resulting in the inability to calibrate and use the product.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS Immunodiagnostic Products AFP Reagent Pack, REF/Product Code 680 0784, Unique Device Identifier No. 10758750000876, IVD; --- Ortho-Clinical Diagnostics Percoed, Bridgend, UK --- Please note: the VITROS Immunodiagnostic Products AFP Calibrators, Unique Device Identifier No. 10758750000913, REF 680 0863, were included in the communication because they are lot-linked to the Reagent Packs and cannot be used with any other lot of reagent. There is no known failure with the Calibrators. For the quantitative measurement of alpha-fetoprotein (AFP) concentrations in human serum using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the management of patients with non-seminomatous testicular cancer.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

Ortho Clinical Diagnostics (OCD) confirmed that specific lots of VITROS Immunodiagnostic Products AFP Reagent Pack product exhibit an increase in the reaction signal over time potentially resulting in the inability to calibrate and use the product.

Which lots are affected?

Lots 6471 (expiry date 19-Jul-2016), 6480 (expiry date 23-Aug-2016)

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656