CLASS II Device recall · Reported 17 Aug 2016 · FDA Enforcement Report

Vitros Immunodiagnostic Products Estradiol Reagent Pack recalled

Ortho-clinical Diagnostics is recalling Vitros Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No, distributed nationwide in the US. The recall was initiated on 14 Jul 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot numbers: 1470 (expiry date 02-Aug-16); 1480 (expiry date 05-Aug-16), 1490 (expiry date 10-Oct-16), 1500 (expiry date 10-Oct-16), 1510 (expiry date 01-Nov-16), 1528 (expiry date 01-Nov-16) 1538 (expiry date 14-Mar-17), 1548 (expiry date 09-Mar-17), 1558 (expiry date 09-Mar-17) and 1568 (expiry date 29-Mar-17)
Quantity recalledDomestic: 1620; Foreign: 7597

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2352-2016
FDA event ID
74764
Recalling firm
Ortho-clinical Diagnostics, Bridgend
Classification
Class II
Status
Terminated
Recall initiated
14 Jul 2016
FDA classified
5 Aug 2016
Reported
17 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Potency & Assay

Timeline

14 Jul 2016Recall initiated by company
5 Aug 2016Classified by FDA+22 days
17 Aug 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ortho Clinical Diagnostics has issued a recall of their VITROS immunodiagnostics Estradiol Reagent assay. The firm was made aware of a potential cross-reactivity between the drug Fulvestrant and theVITROS Immunodiagnostic Estradiol Reagent assay. The firm conducted an internal investigation and was able to confirm biased positive (ranging from 1303-4046%) on Estradiol results, obtained from postmenopausal females taking Fulvestrant.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

VITROS Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No. 10758750005017, REF/Catalog Number/Product Code 855 2630; IVD -- Ortho-Clinical Diagnostics, Pencoed, Bridgend CF35 5PZ, UK --- The VITROS Estradiol test is performed using the VITROS Estradiol Reagent Pack and Calibrators on VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic Systems and VITROS 5600 Integrated Systems.

Where it was distributed

Nationwide, Puerto Rico, and foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, Venezuela.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Ortho Clinical Diagnostics has issued a recall of their VITROS immunodiagnostics Estradiol Reagent assay. The firm was made aware of a potential cross-reactivity between the drug Fulvestrant and theVITROS Immunodiagnostic Estradiol Reagent assay. The firm conducted an internal investigation and was able to confirm biased positive (ranging from 1303-4046%) on Estradiol results, obtained from postmenopausal females taking Fulvestrant.

Which lots are affected?

Lot numbers: 1470 (expiry date 02-Aug-16); 1480 (expiry date 05-Aug-16), 1490 (expiry date 10-Oct-16), 1500 (expiry date 10-Oct-16), 1510 (expiry date 01-Nov-16), 1528 (expiry date 01-Nov-16) 1538 (expiry date 14-Mar-17), 1548 (expiry date 09-Mar-17), 1558 (expiry date 09-Mar-17) and 1568 (expiry date 29-Mar-17)

Where was it sold?

Nationwide, Puerto Rico, and foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656