CLASS II ONGOING Device recall · Reported 25 May 2022 · FDA Enforcement Report

Vitros Immunodiagnostic Products FSH Reagent Pack- IVD recalled

Ortho-Clinical Diagnostics is recalling Vitros Immunodiagnostic Products FSH Reagent Pack- IVD 1931922, distributed nationwide in the US. The recall was initiated on 25 Mar 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUntil further notice, future lots will also be affected. UDI: 10758750000302 Affected Lot Number/ Expiration Date: 3100 01-MAR-2022; 3111 25-MAR-2022; 3120 04-APR-2022; 3130 25-APR-2022; 3140 10-MAY-2022; 3145 10-MAY-2022; 3150 30-MAY-2022; 3160 09-JUN-2022; 3170 22-JUN-2022; 3180 02-AUG-2022; 3190 16-AUG-2022; 3200 06-SEP-2022; 3210 12-SEP-2022.
Quantity recalled2502 total: 2154 US; 348 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1112-2022
FDA event ID
90047
Recalling firm
Ortho-Clinical Diagnostics, Raritan, NJ
Classification
Class II
Status
Ongoing
Recall initiated
25 Mar 2022
FDA classified
18 May 2022
Reported
25 May 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Mar 2022Recall initiated by company
18 May 2022Classified by FDA+54 days
25 May 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922

Where it was distributed

Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom

Nationwide

Questions about this recall

Why was it recalled?

Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.

Which lots are affected?

Until further notice, future lots will also be affected. UDI: 10758750000302 Affected Lot Number/ Expiration Date: 3100 01-MAR-2022; 3111 25-MAR-2022; 3120 04-APR-2022; 3130 25-APR-2022; 3140 10-MAY-2022; 3145 10-MAY-2022; 3150 30-MAY-2022; 3160 09-JUN-2022; 3170 22-JUN-2022; 3180 02-AUG-2022; 3190 16-AUG-2022; 3200 06-SEP-2022; 3210 12-SEP-2022.

Where was it sold?

Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 90047

This product is part of a recall event; the main page for the event is Vitros Immunodiagnostic Products LH Reagent Pack Product recalled.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656