CLASS II ONGOING Device recall · Reported 29 Jan 2025 · FDA Enforcement Report

Vitros 5600 Integrated System recalled by Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics is recalling Vitros 5600 Integrated System, distributed nationwide in the US. The recall was initiated on 19 Dec 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: VITROS 5600 Integrated System (New) Product Code: 6802413 UDI: 10758750002740 VITROS 5600 Integrated System (Certified/Refurbished) Product Code: 6802915 UDI: 10758750002740 Lot Code: VITROS 5600 Integrated System (China) Product Code: 6904001 UDI: 10758750035786 Range of installed serial numbers: 56000015-56004962 Manufacture date range: August 2008 to present Distribution date range; August 2008 to present Total systems currently installed in the US: 905 Total systems currently installed outside the US:2344 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802415 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
Quantity recalled3,249 units (905 US, 2,344 OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1005-2025
FDA event ID
96031
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
19 Dec 2024
FDA classified
21 Jan 2025
Reported
29 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

19 Dec 2024Recall initiated by company
21 Jan 2025Classified by FDA+33 days
29 Jan 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/Refurbished), 6904001 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

Nationwide

Questions about this recall

Is Vitros 5600 Integrated System recalled?

Yes. Ortho-Clinical Diagnostics recalled Vitros 5600 Integrated System on 19 Dec 2024. The recall is Class II and its status is Ongoing. Recall number Z-1005-2025.

Why was it recalled?

The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

Which lots are affected?

Lot Code: VITROS 5600 Integrated System (New) Product Code: 6802413 UDI: 10758750002740 VITROS 5600 Integrated System (Certified/Refurbished) Product Code: 6802915 UDI: 10758750002740 Lot Code: VITROS 5600 Integrated System (China) Product Code: 6904001 UDI: 10758750035786 Range of installed serial numbers: 56000015-56004962 Manufacture date range: August 2008 to present Distribution date range; August 2008 to present Total systems currently installed in the US: 905 Total systems currently installed outside the US:2344 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802415 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.