CLASS II Device recall · Reported 13 Aug 2014 · FDA Enforcement Report

Vitros 5600 Integrated System --- For use recalled

Ortho-Clinical Diagnostics is recalling Vitros 5600 Integrated System --- For use in the in vitro quantitative, semi-quantitative, distributed nationwide in the US. The recall was initiated on 7 Jul 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers J56000110 - J56001954 (J Numbers are analogous to Serial Numbers, for VITROS 5600 systems the J Numbers start with 5600XXXX)
Quantity recalledDomestic: 838, Foreign: 949

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2141-2014
FDA event ID
68790
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Terminated
Recall initiated
7 Jul 2014
FDA classified
1 Aug 2014
Reported
13 Aug 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Jul 2014Recall initiated by company
1 Aug 2014Classified by FDA+25 days
13 Aug 2014Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS 5600 Integrated System --- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents. The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).

Where it was distributed

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

Nationwide Puerto Rico

Questions about this recall

Is Vitros 5600 Integrated System --- For use in the in vitro quantitative, semi-quantitative recalled?

Yes. Ortho-Clinical Diagnostics recalled Vitros 5600 Integrated System --- For use in the in vitro quantitative, semi-quantitative on 7 Jul 2014. The recall is Class II and its status is Terminated. Recall number Z-2141-2014.

Why was it recalled?

Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.

Which lots are affected?

Serial Numbers J56000110 - J56001954 (J Numbers are analogous to Serial Numbers, for VITROS 5600 systems the J Numbers start with 5600XXXX)

Where was it sold?

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.