CLASS II ONGOING Device recall · Reported 13 Mar 2024 · FDA Enforcement Report

Vitros XT 3400 Chemistry System recalled by Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics is recalling Vitros XT 3400 Chemistry System, distributed nationwide in the US. The recall was initiated on 26 Jan 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940
Quantity recalled7 US; 4 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1227-2024
FDA event ID
93925
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
26 Jan 2024
FDA classified
1 Mar 2024
Reported
13 Mar 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

26 Jan 2024Recall initiated by company
1 Mar 2024Classified by FDA+35 days
13 Mar 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Where it was distributed

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

Nationwide ArizonaCaliforniaConnecticutFloridaIowaIdahoIllinoisIndianaKentuckyLouisianaMissouriNebraskaNew YorkOhioOregonSouth CarolinaSouth DakotaTexasWashingtonWisconsin

Questions about this recall

Why was it recalled?

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Which lots are affected?

VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940

Where was it sold?

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 93925

This product is part of a recall event; the main page for the event is Vitros 5600 Integrated System recalled by Ortho-Clinical Diagnostics.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Other Medical Devices recalls

All 13,884