CLASS II ONGOING Device recall · Reported 13 Mar 2024 · FDA Enforcement Report

Vitros XT 7600 Integrated System recalled by Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics is recalling Vitros XT 7600 Integrated System, distributed nationwide in the US. The recall was initiated on 26 Jan 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVITROS System Software Version 3.8.1 UDI-DI 10758750031610 Serial Numbers US: J76000199 J76001383 J76000207 J76001385 J76000353 J76001392 J76000456 J76001404 J76000459 J76001406 J76000460 J76001516 J76000468 J76001518 J76000495 J76001519 J76000502 J76001520 J76000561 J76001581 J76001207 J76001583 J76001230 J76001611 J76001231 J76001620 J76001232 J76001644 J76001235 J76001647 J76001239 J76001660 J76001260 J76001668 J76001263 J76001671 J76001270 J76001674 J76001274 J76001692 J76001331 J76001700 J76001351 J76001748 OUS: J76000118 J76001070 J76000253 J76001074 J76000265 J76001095 J76000274 J76001121 J76000281 J76001130 J76000370 J76001152 J76000447 J76001174 J76000475 J76001328 J76000620 J76001377 J76000855 J76001378 J76000856 J76001396 J76000898 J76001399 J76000901 J76001407 J76000973 J76001408 J76000976 J76001484 J76001001 J76001673 J76001027 J76001761 J76001035
Quantity recalled44 US; 35 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1228-2024
FDA event ID
93925
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
26 Jan 2024
FDA classified
1 Mar 2024
Reported
13 Mar 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

26 Jan 2024Recall initiated by company
1 Mar 2024Classified by FDA+35 days
13 Mar 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS XT 7600 Integrated System, Catalog No. 6844461

Where it was distributed

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

Nationwide ArizonaCaliforniaConnecticutFloridaIowaIdahoIllinoisIndianaKentuckyLouisianaMissouriNebraskaNew YorkOhioOregonSouth CarolinaSouth DakotaTexasWashingtonWisconsin

Questions about this recall

Why was it recalled?

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Which lots are affected?

VITROS System Software Version 3.8.1 UDI-DI 10758750031610 Serial Numbers US: J76000199 J76001383 J76000207 J76001385 J76000353 J76001392 J76000456 J76001404 J76000459 J76001406 J76000460 J76001516 J76000468 J76001518 J76000495 J76001519 J76000502 J76001520 J76000561 J76001581 J76001207 J76001583 J76001230 J76001611 J76001231 J76001620 J76001232 J76001644 J76001235 J76001647 J76001239 J76001660 J76001260 J76001668 J76001263 J76001671 J76001270 J76001674 J76001274 J76001692 J76001331 J76001700 J76001351 J76001748 OUS: J76000118 J76001070 J76000253 J76001074 J76000265 J76001095 J76000274 J76001121 J76000281 J76001130 J76000370 J76001152 J76000447 J76001174 J76000475 J76001328 J76000620 J76001377 J76000855 J76001378 J76000856 J76001396 J76000898 J76001399 J76000901 J76001407 J76000973 J76001408 J76000976 J76001484 J76001001 J76001673 J76001027 J76001761 J76001035

Where was it sold?

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 93925

This product is part of a recall event; the main page for the event is Vitros 5600 Integrated System recalled by Ortho-Clinical Diagnostics.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Other Medical Devices recalls

All 13,884