CLASS II ONGOING Device recall · Reported 21 Aug 2024 · FDA Enforcement Report

Wavewriter Alpha recalled by Boston Scientific

Boston Scientific Neuromodulation is recalling Wavewriter Alpha, distributed nationwide in the US. The recall was initiated on 17 Jul 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365SC12160/8714729985082/100188 - 758920, M365SC12320/8714729985099/100635 - 759404
Quantity recalled77,674

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2617-2024
FDA event ID
94995
Recalling firm
Boston Scientific Neuromodulation, Valencia, CA
Classification
Class II
Status
Ongoing
Recall initiated
17 Jul 2024
FDA classified
14 Aug 2024
Reported
21 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Jul 2024Recall initiated by company
14 Aug 2024Classified by FDA+28 days
21 Aug 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.

Product as listed by the FDA

WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320

Where it was distributed

Worldwide Distribution: US (nationwide) including states of: TN, PA, WA, SC, MN, OH, NC, AK, NY, IL, CA, NJ, KY, IA, MO, UT, FL, IN, LA, AL, MS, MD, WI, GA, VA, TX, AZ, KS, OR, CO, CT, ID, AR, OK, MA, MI, ND, NE, SD, WV, NV, HI, NH, DC, MT, ME, VT, WY, DE, NM. OUS (foreign) countries of: Canada, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Portugal, Japan, Australia, New Zealand, Singapore, Argentina, Brazil, Colombia, Puerto Rico

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.

Which lots are affected?

REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365SC12160/8714729985082/100188 - 758920, M365SC12320/8714729985099/100635 - 759404

Where was it sold?

Worldwide Distribution: US (nationwide) including states of: TN, PA, WA, SC, MN, OH, NC, AK, NY, IL, CA, NJ, KY, IA, MO, UT, FL, IN, LA, AL, MS, MD, WI, GA, VA, TX, AZ, KS, OR, CO, CT, ID, AR, OK, MA, MI, ND, NE, SD, WV, NV, HI, NH, DC, MT, ME, VT, WY, DE, NM. OUS (foreign) countries of: Canada, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Portugal, Japan, Australia, New Zealand, Singapore, Argentina, Brazil, Colombia, Puerto Rico

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Boston Scientific Neuromodulation

All 24

Other Other Medical Devices recalls