CLASS II Device recall · Reported 10 Jun 2020 · FDA Enforcement Report

Xhibit Central recalled by Spacelabs Healthcare

Spacelabs Healthcare is recalling Xhibit Central, distributed nationwide in the US. The recall was initiated on 30 Aug 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel No. 96102 Xhibit Central; Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4; UDI: 10841522100345
Quantity recalled2197 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2261-2020
FDA event ID
85325
Recalling firm
Spacelabs Healthcare, Snoqualmie, WA
Classification
Class II
Status
Terminated
Recall initiated
30 Aug 2019
FDA classified
4 Jun 2020
Reported
10 Jun 2020
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

30 Aug 2019Recall initiated by company
4 Jun 2020Classified by FDA+279 days
10 Jun 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Where it was distributed

Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

Which lots are affected?

Model No. 96102 Xhibit Central; Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4; UDI: 10841522100345

Where was it sold?

Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.