CLASS II Device recall · Reported 9 Nov 2016 · FDA Enforcement Report
Xhibit Central Station recalled by Spacelabs Healthcare
Spacelabs Healthcare is recalling Xhibit Central Station, distributed in 28 states. The recall was initiated on 18 Oct 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0332-2017
- FDA event ID
- 75477
- Recalling firm
- Spacelabs Healthcare, Snoqualmie, WA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 18 Oct 2016
- FDA classified
- 2 Nov 2016
- Reported
- 9 Nov 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 18 Oct 2016 | Recall initiated by company | |
| 2 Nov 2016 | Classified by FDA | +15 days |
| 9 Nov 2016 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.
Where it was distributed
Worldwide Distribution -- USA, to AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MI, MN, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, WY, and Puerto Rico; and, to the countries of AUSTRALIA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, MEXICO, NETHERLANDS, PANAMA, SAUDI ARABIA, SWITZERLAND, TAIWAN, and UNITED KINGDOM.
AlabamaCaliforniaConnecticutFloridaGeorgiaIowaIdahoIllinoisIndianaKentuckyMichiganMinnesotaMississippiNorth CarolinaNebraskaNew JerseyNew YorkOhioOklahomaPennsylvania
Questions about this recall
Why was it recalled?
The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.
Which lots are affected?
Software version 1.1.5, Affected Serial Numbers in the US ************************************* ADDITIONAL FIVE UNITS ADDED: SLA0216F01796; SLA0216F01808; SLA1215F01638; SLA1215F01645; SLA1215F01652; SLA0114F00445, SLA0114F00446, SLA0114F00447, SLA0115F00917, SLA0115F00918, SLA0115F00919, SLA0115F00921, SLA0115F00923, SLA0115F00924, SLA0115F00927, SLA0115F00928, SLA0115F00934, SLA0115F00955, SLA0115F00961, SLA0115F00968, SLA0115F00969, SLA0115F00972, SLA0115F00976, SLA0116F01718, SLA0116F01720, SLA0116F01722, SLA0214F00489, SLA0216F01807, SLA0216F01809, SLA0216F01813, SLA0216F01821, SLA0216F01824, SLA0216F01837, SLA0216F01840, SLA0216F01844, SLA0314F00534, SLA0314F00535, SLA0314F00540, SLA0314F00543, SLA0314F00547, SLA0314F00548, SLA0314F00549, SLA0314F00552, SLA0314F00561, SLA0314F00567, SLA0414F00568, SLA0414F00585, SLA0416F01850, SLA0416F01853, SLA0416F01857, SLA0416F01860, SLA0416F01875, SLA0416F01876, SLA0416F01883, SLA0416F01885, SLA0416F01889, SLA0416F01895, SLA0416F01923, SLA0516F01969, SLA0516F01970, SLA0516F01972, SLA0516F01974, SLA0516F01975, SLA0516F01976, SLA0516F01977, SLA0516F01978, SLA0516F01981, SLA0516F01982, SLA0516F01985, SLA0516F01986, SLA0516F01987, SLA0516F01992, SLA0516F02002, SLA0516F02005, SLA0614F00631, SLA0614F00654, SLA0615F01188, SLA0615F01194, SLA0615F01197, SLA0615F01198, SLA0615F01199, SLA0615F01213, SLA0615F01237, SLA0615F01238, SLA0615F01239, SLA0615F01240, SLA0615F01244, SLA0615F01248, SLA0615F01261, SLA0714F00678, SLA0714F00688, SLA0714F00694, SLA0714F00702, SLA0714F00703, SLA0715F01274, SLA0715F01310, SLA0715F01314, SLA0715F01325, SLA0715F01331, SLA0715F01334, SLA0715F01346, SLA0715F01347, SLA0716F02009, SLA0716F02011, SLA0716F02013, SLA0716F02015, SLA0716F02016, SLA0716F02017, SLA0716F02020, SLA0716F02021, SLA0716F02022, SLA0716F02028, SLA0716F02034, SLA0716F02042, SLA0716F02043, SLA0716F02044, SLA0716F02045, SLA0814F00713, SLA0815F01371, SLA0815F01372, SLA0815F01373, SLA0815F01374, SLA0815F01375, SLA0815F01376, SLA0815F01379, SLA0815F01380, SLA0815F01381, SLA0815F01382, SLA0815F01383, SLA0815F01384, SLA0815F01386, SLA0815F01387, SLA0815F01388, SLA0815F01390, SLA0815F01391, SLA0815F01395, SLA0815F01397, SLA0815F01400, SLA0815F01402, SLA0815F01403, SLA0815F01404, SLA0815F01405, SLA0815F01406, SLA0815F01407, SLA0815F01408, SLA0815F01409, SLA0815F01412, SLA0815F01415, SLA0815F01418, SLA0815F01419, SLA0815F01424, SLA0816F02048, SLA0816F02050, SLA0816F02051, SLA0816F02052, SLA0816F02053, SLA0816F02054, SLA0816F02055, SLA0816F02056, SLA0816F02057, SLA0816F02058, SLA0816F02059, SLA0816F02060, SLA0816F02061, SLA0816F02062, SLA0816F02063, SLA0816F02065, SLA0816F02066, SLA0816F02067, SLA0816F02074, SLA0816F02075, SLA0816F02076, SLA0816F02077, SLA0816F02078, SLA0816F02081, SLA0816F02082, SLA0816F02083, SLA0816F02085, SLA0816F02087, SLA0816F02089, SLA0816F02091, SLA0816F02092, SLA0816F02093, SLA0816F02094, SLA0816F02095, SLA0816F02096, SLA0816F02097, SLA0816F02098, SLA0816F02099, SLA0816F02100, SLA0816F02101, SLA0816F02102, SLA0816F02103, SLA0816F02104, SLA0816F02105, SLA0816F02106, SLA0816F02107, SLA0915F01439, SLA0915F01442, SLA0915F01447, SLA0915F01451, SLA0915F01452, SLA0915F01457, SLA0915F01459, SLA0915F01462, SLA0915F01468, SLA0915F01474, SLA0915F01493, SLA0915F01496, SLA0915F01497, SLA0915F01498, SLA0915F01499, SLA0915F01501, SLA0915F01503, SLA0915F01506, SLA0915F01507, SLA0915F01508, SLA0915F01510, SLA0915F01511, SLA0915F01512, SLA0915F01513, SLA0915F01514, SLA0916F02108, SLA0916F02110, SLA0916F02111, SLA0916F02112, SLA0916F02114, SLA0916F02115, SLA0916F02117, SLA0916F02121, SLA0916F02122, SLA0916F02123, SLA0916F02126, SLA0916F02127, SLA0916F02130, SLA0916F02131, SLA0916F02132, SLA0916F02133, SLA0916F02134, SLA0916F02136, SLA0916F02137, SLA0916F02138, SLA0916F02139, SLA0916F02141, SLA0916F02142, SLA0916F02143, SLA0916F02145, SLA0916F02146, SLA0916F02147, SLA0916F02148, SLA0916F02149, SLA0916F02150, SLA0916F02153, SLA0916F02154, SLA0916F02155, SLA0916F02156, SLA0916F02158, SLA0916F02159, SLA0916F02160, SLA0916F02162, SLA0916F02163, SLA0916F02165, SLA0916F02166, SLA1015F01525, SLA1015F01528, SLA1015F01530, SLA1015F01540, SLA1015F01551, SLA1015F01560, SLA1114F00795, SLA1114F00820, SLA1115F01578, SLA1115F01579, SLA1115F01582, SLA1115F01584, SLA1115F01593, SLA1115F01594, SLA1115F01595, SLA1115F01596, SLA1115F01597, SLA1115F01599, SLA1115F01602, SLA1115F01603, SLA1115F01608, SLA1115F01609, SLA1214F00858, SLA1214F00871, SLA1214F00876, SLA1214F00881, SLA1214F00885, SLA1214F00886, and SLA1215F01632. ***********Affected Serial Numbers OUTSIDE the US ************************** SLA0114F00430, SLA0114F00437, SLA0114F00438, SLA0114F00439, SLA0114F00440, SLA0114F00443, SLA0115F00975, SLA0116F01724, SLA0314F00528, SLA0314F00541, SLA0414F00596, SLA0416F01872, SLA0416F01877, SLA0416F01880, SLA0612F00235, SLA0715F01270, SLA0715F01271, SLA0715F01273, SLA0715F01283, SLA0715F01335, SLA0716F02023, SLA0716F02025, SLA0716F02029, SLA0716F02031, SLA0716F02035, SLA0716F02036, SLA0716F02037, SLA0716F02040, SLA0716F02041, SLA0716F02047, SLA0815F01431, SLA0816F02049, SLA0816F02064, SLA0816F02069, SLA0816F02070, SLA0816F02071, SLA0816F02072, SLA0816F02073, SLA0816F02079, SLA0816F02080, SLA0816F02084, SLA0816F02086, SLA0916F02109, SLA0916F02118, SLA0916F02119, SLA0916F02120, SLA0916F02128, SLA0916F02129, SLA0916F02140, SLA0916F02161, SLA1014F00776, SLA1113F00348, SLA1113F00368, SLA1115F01604, SLA1115F01605, and SLA1213F00389.
Where was it sold?
Worldwide Distribution -- USA, to AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MI, MN, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, WY, and Puerto Rico; and, to the countries of AUSTRALIA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, MEXICO, NETHERLANDS, PANAMA, SAUDI ARABIA, SWITZERLAND, TAIWAN, and UNITED KINGDOM.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Spacelabs Healthcare
All 47| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Xhibit Central Station recalled by Spacelabs Healthcare | Spacelabs Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Spacelabs Ultraview SL Command Module SL Command Module recalled | Spacelabs Healthcare | Packaging Defects | Nationwide |
| Device | CLASS II | Multi-parameter Command Module recalled by Spacelabs Healthcare | Spacelabs Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Sentinel V11 recalled by Spacelabs Healthcare | Spacelabs Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Branding is not applied to this product DVI display cables recalled | Spacelabs Healthcare | Device Malfunction | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |