CLASS II ONGOING Device recall · Reported 15 Jan 2025 · FDA Enforcement Report

Xhibit Telemetry Receiver recalled by Spacelabs Healthcare

Spacelabs Healthcare is recalling Xhibit Telemetry Receiver, distributed in 27 states. The recall was initiated on 25 Nov 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246
Quantity recalled434

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0810-2025
FDA event ID
95910
Recalling firm
Spacelabs Healthcare, Snoqualmie, WA
Classification
Class II
Status
Ongoing
Recall initiated
25 Nov 2024
FDA classified
7 Jan 2025
Reported
15 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

25 Nov 2024Recall initiated by company
7 Jan 2025Classified by FDA+43 days
15 Jan 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

Where it was distributed

US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.

AlabamaArkansasColoradoFloridaGeorgiaIowaIdahoIllinoisKansasLouisianaMassachusettsMaineMinnesotaMissouriMississippiNew MexicoNew YorkOhioOklahomaOregon

Questions about this recall

Why was it recalled?

Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.

Which lots are affected?

Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246

Where was it sold?

US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Spacelabs Healthcare

All 47

Other Patient Monitors recalls

All 1,022