CLASS II ONGOING Device recall · Reported 1 Aug 2012 · FDA Enforcement Report

ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp recalled

Stryker Howmedica Osteonics is recalling ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp, distributed nationwide in the US. The recall was initiated on 30 Apr 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes510 K092406 all lots (4845-4XXX and NLS/NLV)
Quantity recalled9003 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2089-2012
FDA event ID
62456
Recalling firm
Stryker Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Ongoing
Recall initiated
30 Apr 2012
FDA classified
26 Jul 2012
Reported
1 Aug 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

30 Apr 2012Recall initiated by company
26 Jul 2012Classified by FDA+87 days
1 Aug 2012Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an

Product as listed by the FDA

ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill Industrial Estate, Carrigtwohill County Cork, Ireland. The ABG II Modular range of implants is constituted of 8 right stems and 8 left stems. The necks are made of GADS Vitallium, a proprietary Co Cr alloy that Stryker has developed. GADS stands for Gas Atomized Dispersion Strengthened. IT is a Co Cr alloy that is a corrosion resistant and demonstrates qualities of improved fatigue strength (with respect to regular Co Cr alloy). Neck Range - The ABG II modular neck range is composed of 10 reversible neck implants. There are different options in terms of length, version and neck angle. The AG II Modular necks come in two different lengths: 28 mm (short) and 36 mm (long). The necks are offered in two version angles ; 0 and 7. The 0 and 7 necks are available in 125, 130 and 135 neck angles.

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an

Which lots are affected?

510 K092406 all lots (4845-4XXX and NLS/NLV)

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 62456
ReportedClassRecallCompanyReasonWhere
CLASS IIRejuvenate Modular Stems Stryker Ireland Carrigtwohill recalledStryker Howmedica OsteonicsOtherNationwide

More recalls from Stryker Howmedica Osteonics

All 86

Other Hip & Knee Implants recalls

All 3,111