CLASS II Device recall · Reported 5 Apr 2017 · FDA Enforcement Report

AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is recalled

AGFA Healthcare is recalling AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing, distributed nationwide in the US. The recall was initiated on 20 Jan 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesA5410000053, A5410000054, A5411000439, A5411000446, A5411000447, A5411000448, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5410000100, A5410000107, A5410000134, A5411000184, A5410000039, A5411000140, A5411000142, A5411000314, A5411000342, A5411000347, A5411000349, A5410000007, A5411000017, A5411000315, A5411000331, A5411000332, A5411000463, A5411000141, A5411000144, A5411000539, A5411000541, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000019, A5411000400, A5411000401, A5411000540, A5411000047, A5411000048, A5411000053, A5410000041, A5410000042, A5410000046, A5411000041, A5411000079, A5411000024, A5411000032, A5411000083, A5411000038, A5411000132, A5411000571, A5411000573, A5411000050, A5411000088, A5411000247, A5411000070, A5411000078, A5411000082, A5411000476, A5411000478, A5411000480, A5411000482, A5411000489, A5411000490, A5411000563, A5411000582, A5411000154, A5411000156, A5411000404, A5411000500, A5411000501, A5411000589, A5411000591, A5411000583, A5411000236, A5411000155, A5411000151, A5411000191, A5411000192, A5411000194, A5411000193, A5411000249, A5411000052, A5411000185, A5411000237, A5411000030, A5411000397, A5411000587, A5411000499, A5411000503, A5411000504, A5411000574, A5411000584, A5411000348, A5411000193, A5411000268, A5411000498, A5411000502, A5411000509, A5411000403, A5411000153, A5411000233, A5411000235, A5411000318, A5411000398, A5411000538, A5411000371, A5410000043, A5411000181, A5411000158, A5411000513, A5411000514, A5411000152, A5411000548, A5411000553, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000468, A5411000469, A5411000470, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000336, A5411000533, A5411000537, A5411000536, A5411000159, A5410000038, A5411000148, A5411000149, A5411000134, A5411000405, A5411000120, A5411000125, A5411000372, A5411000147, A5411000588, A5411000039, A5411000117, A5411000121, A5411000333, A5411000334, A5411000377, A5411000040, A5411000240, A5411000593, A5411000594, A5411000085, A5411000419, A5411000097, A5411000281, A5411000278, A5411000279, A5411000280, A5411000492, A5411000282, A5411000245, A5411000277, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000399, A5411000491, A5411000493. A5411000524, A5411000497, A5411000534, A5411000526, A5411000535, A5411000122, A5411000575, A5411000576, A5411000577, A5411000495, A5411000578, A5410000044, A5411000037, A5411000241, A5411000246, A5411000108, A5411000110, A5411000114, A5411000312, A5411000074, A5411000080, A5411000522, A5411000033, A5411000139, A5411000543, A5411000182, A5411000190, A5411000021, A5411000374, A5411000218, A5411000572, A5411000084, A5411000098, A5411000238, A5411000116, A5411000408, A5411000409, A5410000042, A5410000023, A5411000544, A5411000580, A5411000581, A5411000545, A5411000546, A5411000547, A5411000579, A5411000337, A5411000338, A5411000340, A5411000341, A5411000475, A5411000549, A5411000373, A5411000420, A5410000133, A5410000117, A5410000115, A5410000079, A5410000131, A5410000066, A5410000132, A5410000114, A5410000082, A5410000116, A5411000025, A5411000523, A5411000525, A5411000590, A5411000592, A5411000263, A5411000309, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000363, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000034, A5411000081, A5411000146, A5411000092, A5411000093, A5411000407, A5411000026, A5411000068, A5411000367, A5411000368, A5411000465, A5411000060, A5411000131, A5411000018, A5411000100, A5411000515, A5411000313, A5411000284, A5411000076, A5411000135, A5411000180, A5411000343, A5411000285, A5411000311, A5411000375, A5411000062, A5411000136, A5411000179, A5411000178, A5411000421, A5411000427, A5411000376
Quantity recalled308

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1698-2017
FDA event ID
76806
Recalling firm
AGFA Healthcare, Greenville, SC
Classification
Class II
Status
Terminated
Recall initiated
20 Jan 2016
FDA classified
29 Mar 2017
Reported
5 Apr 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

20 Jan 2016Recall initiated by company
29 Mar 2017Classified by FDA+434 days
5 Apr 2017Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.

Product as listed by the FDA

AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis

Where it was distributed

Worldwide Distribution - US including AL, AR, CA, CO, DC, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NY, NC, OH, OK, OR, RI, SC, TN, TX, VA, WV, WI, WY, and Internationally to Canada.

Nationwide AlabamaArkansasCaliforniaColoradoDistrict of ColumbiaFloridaGeorgiaIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiNorth Carolina

Questions about this recall

Is AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing recalled?

Yes. AGFA Healthcare recalled AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing on 20 Jan 2016. The recall is Class II and its status is Terminated. Recall number Z-1698-2017.

Why was it recalled?

The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.

Which lots are affected?

A5410000053, A5410000054, A5411000439, A5411000446, A5411000447, A5411000448, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5410000100, A5410000107, A5410000134, A5411000184, A5410000039, A5411000140, A5411000142, A5411000314, A5411000342, A5411000347, A5411000349, A5410000007, A5411000017, A5411000315, A5411000331, A5411000332, A5411000463, A5411000141, A5411000144, A5411000539, A5411000541, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000019, A5411000400, A5411000401, A5411000540, A5411000047, A5411000048, A5411000053, A5410000041, A5410000042, A5410000046, A5411000041, A5411000079, A5411000024, A5411000032, A5411000083, A5411000038, A5411000132, A5411000571, A5411000573, A5411000050, A5411000088, A5411000247, A5411000070, A5411000078, A5411000082, A5411000476, A5411000478, A5411000480, A5411000482, A5411000489, A5411000490, A5411000563, A5411000582, A5411000154, A5411000156, A5411000404, A5411000500, A5411000501, A5411000589, A5411000591, A5411000583, A5411000236, A5411000155, A5411000151, A5411000191, A5411000192, A5411000194, A5411000193, A5411000249, A5411000052, A5411000185, A5411000237, A5411000030, A5411000397, A5411000587, A5411000499, A5411000503, A5411000504, A5411000574, A5411000584, A5411000348, A5411000193, A5411000268, A5411000498, A5411000502, A5411000509, A5411000403, A5411000153, A5411000233, A5411000235, A5411000318, A5411000398, A5411000538, A5411000371, A5410000043, A5411000181, A5411000158, A5411000513, A5411000514, A5411000152, A5411000548, A5411000553, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000468, A5411000469, A5411000470, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000336, A5411000533, A5411000537, A5411000536, A5411000159, A5410000038, A5411000148, A5411000149, A5411000134, A5411000405, A5411000120, A5411000125, A5411000372, A5411000147, A5411000588, A5411000039, A5411000117, A5411000121, A5411000333, A5411000334, A5411000377, A5411000040, A5411000240, A5411000593, A5411000594, A5411000085, A5411000419, A5411000097, A5411000281, A5411000278, A5411000279, A5411000280, A5411000492, A5411000282, A5411000245, A5411000277, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000399, A5411000491, A5411000493. A5411000524, A5411000497, A5411000534, A5411000526, A5411000535, A5411000122, A5411000575, A5411000576, A5411000577, A5411000495, A5411000578, A5410000044, A5411000037, A5411000241, A5411000246, A5411000108, A5411000110, A5411000114, A5411000312, A5411000074, A5411000080, A5411000522, A5411000033, A5411000139, A5411000543, A5411000182, A5411000190, A5411000021, A5411000374, A5411000218, A5411000572, A5411000084, A5411000098, A5411000238, A5411000116, A5411000408, A5411000409, A5410000042, A5410000023, A5411000544, A5411000580, A5411000581, A5411000545, A5411000546, A5411000547, A5411000579, A5411000337, A5411000338, A5411000340, A5411000341, A5411000475, A5411000549, A5411000373, A5411000420, A5410000133, A5410000117, A5410000115, A5410000079, A5410000131, A5410000066, A5410000132, A5410000114, A5410000082, A5410000116, A5411000025, A5411000523, A5411000525, A5411000590, A5411000592, A5411000263, A5411000309, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000363, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000034, A5411000081, A5411000146, A5411000092, A5411000093, A5411000407, A5411000026, A5411000068, A5411000367, A5411000368, A5411000465, A5411000060, A5411000131, A5411000018, A5411000100, A5411000515, A5411000313, A5411000284, A5411000076, A5411000135, A5411000180, A5411000343, A5411000285, A5411000311, A5411000375, A5411000062, A5411000136, A5411000179, A5411000178, A5411000421, A5411000427, A5411000376

Where was it sold?

Worldwide Distribution - US including AL, AR, CA, CO, DC, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NY, NC, OH, OK, OR, RI, SC, TN, TX, VA, WV, WI, WY, and Internationally to Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.