CLASS II Device recall · Reported 21 Jan 2015 · FDA Enforcement Report
Philips Expression MR200 MRI Patient Monitoring System recalled
Invivo is recalling Philips Expression MR200 MRI Patient Monitoring System, distributed in 11 states. The recall was initiated on 22 Oct 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0958-2015
- FDA event ID
- 69587
- Recalling firm
- Invivo, Orlando, FL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 22 Oct 2014
- FDA classified
- 13 Jan 2015
- Reported
- 21 Jan 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Other Reasons
- Device type
- Imaging Equipment
Timeline
| 22 Oct 2014 | Recall initiated by company | |
| 13 Jan 2015 | Classified by FDA | +83 days |
| 21 Jan 2015 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock.
Product as listed by the FDA
Philips Expression MR200 MRI Patient Monitoring System.
Where it was distributed
CA, FL, IN, KY, MN, NC, NJ, NY, OK, SD and TX Netherlands, Poland and United Kingdom.
CaliforniaFloridaIndianaKentuckyMinnesotaNorth CarolinaNew JerseyNew YorkOklahomaSouth DakotaTexas
Questions about this recall
Why was it recalled?
The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock.
Which lots are affected?
Model No. 866120; Lot/ Serial Numbers: US33600065, US33600066, US33600075, US33600077, US33600078, US33600079; US33600080, US33600081, US33600082, US33600083, US33600084, US33600086, US33600087, US33600088, US33600089, US33600090, US33600091, US33600092, US33600093, US33600094.
Where was it sold?
CA, FL, IN, KY, MN, NC, NJ, NY, OK, SD and TX Netherlands, Poland and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Invivo
All 18| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | MR Coils The MR Coil is intended to be used recalled by Invivo | Invivo | Other | Nationwide |
| Device | CLASS II | Monitor for the Functional Imaging System recalled by Invivo | Invivo | Other | 17 states |
| Device | CLASS II | Xper Flex Cardio Physiomonitoring System recalled by Invivo | Invivo | Device Malfunction | FL |
| Device | CLASS II | Xper Flex Cardio Physiomonitoring System recalled by Invivo | Invivo | Labeling Errors | FL |
| Device | CLASS II | Xper Flex Cardio Physiomonitoring System The Xper Flex recalled | Invivo | Temperature Abuse | Nationwide |
Other Imaging Equipment recalls
All 2,341| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips EPIQ Elite Diagnostic Ultrasound System recalled | Philips Ultrasound | Labeling Errors | Nationwide |
| Device | CLASS II | GmbH Spectralis with Flex Module recalled by Heidelberg Engineering | Heidelberg Engineering | Potency & Assay | Nationwide |
| Device | CLASS II | Core Services Hub V8, a set of backend services including recalled | GE Healthcare | Other | Nationwide |
| Device | CLASS II | Centricity PACS V2, V3, V4, V6, V7 recalled by GE Healthcare | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS I | Haylard Univ Imaging Biopsy Trucustom kits recalled by Avid Medical | Avid Medical | Sterility | 4 states |