CLASS II ONGOING Device recall · Reported 5 Apr 2023 · FDA Enforcement Report

SMR TT Hybrid glenoid Std - Peg S recalled by Limacorporate S.p.A

Limacorporate is recalling SMR TT Hybrid glenoid Std - Peg S, distributed in Missouri. The recall was initiated on 21 Feb 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI/Lot/Expiration: 08033390121798/2206565/2027-04-30
Quantity recalled1

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1317-2023
FDA event ID
91869
Recalling firm
Limacorporate, San Daniele Del Friuli, N/A
Classification
Class II
Status
Ongoing
Recall initiated
21 Feb 2023
FDA classified
30 Mar 2023
Reported
5 Apr 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

21 Feb 2023Recall initiated by company
30 Mar 2023Classified by FDA+37 days
5 Apr 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Shoulder glenoid fossa prosthesis was manufactured without tantalum markers so in the absence of the markers, on already implanted devices, a correct radiographic assessment of the hybrid glenoid position can be performed by the evaluation of the joint space, central TT peg position and cement mantel observation.

Product as listed by the FDA

SMR TT Hybrid glenoid Std - Peg S, REF: 1379.59.210

Where it was distributed

US: MO

Missouri

Questions about this recall

Why was it recalled?

Shoulder glenoid fossa prosthesis was manufactured without tantalum markers so in the absence of the markers, on already implanted devices, a correct radiographic assessment of the hybrid glenoid position can be performed by the evaluation of the joint space, central TT peg position and cement mantel observation.

Which lots are affected?

UDI-DI/Lot/Expiration: 08033390121798/2206565/2027-04-30

Where was it sold?

US: MO

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Limacorporate

All 16

Other Other Medical Devices recalls

All 13,986