CLASS II Device recall · Reported 20 Nov 2013 · FDA Enforcement Report

Titan 3T MRI System, MRT-3010A/5 recalled by Toshiba American Medical

Toshiba American Medical Systems is recalling Titan 3T MRI System, MRT-3010A/5, distributed nationwide in the US. The recall was initiated on 28 Oct 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002.
Quantity recalled18 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0201-2014
FDA event ID
66725
Recalling firm
Toshiba American Medical Systems, Tustin, CA
Classification
Class II
Status
Terminated
Recall initiated
28 Oct 2013
FDA classified
12 Nov 2013
Reported
20 Nov 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

28 Oct 2013Recall initiated by company
12 Nov 2013Classified by FDA+15 days
20 Nov 2013Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.

Product as listed by the FDA

Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.

Where it was distributed

USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.

Nationwide CaliforniaFloridaGeorgiaIowaIndianaLouisianaNevadaNew YorkOregonPennsylvaniaTexasWisconsin

Questions about this recall

Why was it recalled?

Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.

Which lots are affected?

Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002.

Where was it sold?

USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Toshiba American Medical Systems

All 57

Other Imaging Equipment recalls

All 2,336